EchoTip ClearCore G60992Recalls & reported problems

Cook Ireland Ltd · Echotip ClearCore EUS Biopsy Needle · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002609921), received between Sep 18, 2025 and Jan 19, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Accessory Incompatible1
  • Loss of or Failure to Bond1
  • Material Twisted/Bent1
  • Off-Label Use1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2026Loss of or Failure to Bond

    A COMBINED INITIAL SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 23 DEC 2025. DESCRIPTION OF EVENT: THE CUSTOMER OPENED THE NEEDLE; THEY SAW THE LITTLE PLASTIC PART AT THE END OF THE NEEDLE WAS BROKEN (WHERE THE RECOIL STYLET GOES IN). THEY OPENED ANOTHER NEEDLE TO FINISH PROCEDURE. ANSWER TO ADDITIONAL QUESTIONS RECEIVED ON 04-NOV 25. 1. CAN PHOTOS OF THE DAMAGE BE PROVIDED? I SENT THE DEVICE IN TO COOK. 2. PLEASE DESCRIBE THE DAMAGE IN DETAIL: UPON OPENING THEY NOTICED THE PLASTIC PIECE AT THE BACK END OF THE NEEDLE WHERE THE RECOIL STYLET GOES WAS BROKEN 3. WAS THE DEVICE DAMAGE IDENTIFIED: A) PRIOR TO OPENING THE PACKAGE. -WHEN OPENING THE PACKAGE, THEY SAW IT WAS BROKEN. 4. WAS THE FUNCTIONALITY OF THE DEVICE TESTED PRIOR TO NOTICING THE DAMAGE? · IF YES, PLEASE PROVIDE ADDITIONAL DETAILS (INCLUDING OTHER DEVICES THAT MAY HAVE ENGAGED WITH THE COOK DEVICE). NO. 5. WERE ANY PREPARATORY STEPS PERFORMED TO THE DEVICE (PER INSTRUCTIONS FOR USE, IF APPLICABLE) PRIOR TO NOTICING THE DAMAGE? · IF SO, PLEASE DESCRIBE HOW THE DEVICE WAS PREPARED: NO 6. WAS THE DEVICE STORED ACCORDING TO THE STORAGE CONDITIONS ON THE LABEL (IF APPLICABLE)? · IF NO,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109462 ↗
  • MalfunctionOct 28, 2025Retraction Problem; Material Twisted/Bent; Retraction Problem; Material Twisted/Bent

    THE NEEDLE TIP WOULD NOT GO FULLY BACK INTO THE SHEATH. THEY GOT ANOTHER DEVICE TO FINISH THE PROCEDURE. 1. PLEASE DESCRIBE THE LOCATION IN THE BODY FOR THE INTENDED TARGET SITE (PANCREAS, STOMACH, LUNGS, ETC.) (IF LUNGS, WHICH LYMPH NODE WAS BEING TARGETED? E.G., 2R, 2L, 4R, AO, AR, 11RI, 11S, ETC.) PANCREATIC HEAD 2. WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? N/A, YES, NO YES 3. WAS PUNCTURE OF THE TARGETED SITE DIFFICULT? NO 4. WHAT IS THE SCOPE MANUFACTURER AND MODEL NUMBER THAT WAS USED? OLYMPUS 180 5. WAS THE SCOPE RECENTLY SERVICE/REPAIRED? NO 6. WAS THE DEVICE USED IN A TORTUOUS POSITION? YES 7. WAS FORCE REQUIRED TO REMOVE THE DEVICE? N/A, YES, NO 8. WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DELIVERY SYSTEM (E.G., KINKS) PRIOR TORETURN? (IF YES, PLEASE SPECIFY WHAT ADDITIONAL DEFECT(S) WERE OBSERVED AND WHERE ON THE DEVICE THIS WAS OBSERVED) NO 9. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES AS A RESULT OF THIS EVENT? N/A, YES, NO10. IF ADDITIONAL INTERVENTION WAS PERFORMED, WAS IT DURING THE SAME PROCEDURE AS THE DEVICE FAILURE OR WAS IT SCHEDULED FOR ANOTHER DAY? N/A 11. WAS THERE DIFFICULTY LOCKING THE SHEATH (OR NEEDLE) IN PLACE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23401906 ↗
  • MalfunctionSep 18, 2025Accessory Incompatible; Off-Label Use; Accessory Incompatible; Off-Label Use

    WE USE THIS PRODUCT FOR COELIC PLEXUS BLOCK, WHEN WE ATTACHED THE 10ML SYRINGE THE NEEDLE CONNECTOR IS NOT FIX PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23085723 ↗

Other Endoscopic ultrasound system, gastroenterology-urology devices

All Endoscopic ultrasound system, gastroenterology-urology devices →