Skip to content

Intervertebral Body Fusion Device 10-P3215-13PRecalls & reported problems

Precision Spine, Inc. · Vault ALIF Peek Plate 32mm x 13mm 15 deg Lordosis · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019908528), received between May 10, 2021 and May 10, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Mechanical Problem2
  • Unstable2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 10, 2021Mechanical Problem; Unstable; Detachment of Device or Device Component

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6), 2021, IN EQUADOR, UTILIZING THE VAULT ALIF SYSTEM. DURING THE PROCEDURE TWO (2) OF THE VAULT ALIF PLATES DID NOT ENGAGE ONTO THE CAGE. ALSO, ONE (1) CAGE AND PLATE BECAME UNSTABLE BECAUSE THE CONNECTOR SCREW CAME OFF. THE DOCTOR WAS ABLE TO COMPLETE THE CASE USING OTHER VAULT ALIF CAGES FROM THE SET. THERE WAS NO PATIENT INJURY REPORTED BUT THERE WAS A FORTY-FIVE (45) MINUTE DELAY OF THE PROCEDURE BECAUSE OF THE REPORTED ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 11801825 ↗
  • InjuryMay 10, 2021Mechanical Problem; Unstable; Detachment of Device or Device Component

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON FEBRUARY 5, 2021, IN EQUADOR, UTILIZING THE VAULT ALIF SYSTEM. DURING THE PROCEDURE TWO (2) OF THE VAULT ALIF PLATES DID NOT ENGAGE ONTO THE CAGE. ALSO, ONE (1) CAGE AND PLATE BECAME UNSTABLE BECAUSE THE CONNECTOR SCREW CAME OFF. THE DOCTOR WAS ABLE TO COMPLETE THE CASE USING OTHER VAULT ALIF CAGES FROM THE SET. THERE WAS NO PATIENT INJURY REPORTED BUT THERE WAS A FORTY-FIVE (45) MINUTE DELAY OF THE PROCEDURE BECAUSE OF THE REPORTED ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 11801794 ↗

Other Intervertebral Fusion Device With Integrated Fixation, Lumbar devices

All Intervertebral Fusion Device With Integrated Fixation, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page