Skip to content

MectaLIF Anterior 03.30.302Recalls & reported problems

Medacta International SA · MectaLIF Ant - Lag Plate Flush H14 mm · Class 2 · Implant

All models of MectaLIF Anterior (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630345732309), received between Oct 30, 2025 and Oct 30, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 8, 2023Class IICompleted

    There is a potential for breakage of the small screw used to affix the anti-backout plate.

    Z-1402-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2025Fracture; Fracture

    DURING PRIMARY SPINAL SURGERY AT L4-5, THE FIXED ANGLE AWL BROKE OFF IN THE BONE, THE SURGEON PULLED IT OUT AND USED A DIFFERENT FIXED ANGLE AWL TO MAKE THE HOLE FOR THE SCREW. SURGICAL STEPS PERFORMED: DISCECTOMY, TRIAL, MECTALIF IMPLANT, AWL FOR SCREW HOLES, SCREWS INSERTION AND THE COVER PLATE PLACEMENT. NO TAPPING/DRILLING PERFORMED. DURING CAUDAL SCREW INSERTION, THE SCREWDRIVER BROKE, AND A BACKUP SCREWDRIVER WAS REQUIRED. ADDITIONALLY, WHILE ATTACHING THE ANTI-BACKOUT LAG PLATE, THE ANTI-BACKOUT SCREW ON THE COVER PLATE BROKE OFF WHEN BEING TIGHTENED BY THE TORQUE LIMITER. IT WAS NOTED THAT THE BONE WAS SCLEROTIC, AND THE SURGEON DETERMINED THERE WAS NO RISK OF SCREW BACKOUT. THE BROKEN PORTION OF THE SCREW REMAINED IN THE ANTERIOR PLATE, PREVENTING THE PLACEMENT OF ANOTHER COVER PLATE. THE SURGERY WAS COMPLETED WITHOUT THE COVER PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23424083 ↗

Other Intervertebral Fusion Device With Integrated Fixation, Lumbar devices

All Intervertebral Fusion Device With Integrated Fixation, Lumbar devices →

Sources and updates