Reform Pedicle Screw System 39-CH-0008Recalls & reported problems
Precision Spine, Inc. · Offset Ratcheting Torque Handle · Class 1
All models of Reform Pedicle Screw System (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840019928212), received between Jan 23, 2024 and Jan 23, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2024Fracture; Fracture
IT WAS REPORTED THAT AN L4-L5 POSTERIOR LUMBAR FUSION WAS PERFORMED ON (B)(6) 2023, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. IN THE FINAL STEP, THE DOCTOR RETRIEVED THE STANDARD TORQUE HANDLE (39-CH-0008) WITH A STANDARD SHAFT (39-RD-0060). THE INITIAL LEFT AND RIGHT SIDE SCREWS WERE (39-SB-6540 & 6545) WERE SOLIDLY PLACED IN ALL FOUR (4) PEDICLES. FOUR (4) MODULAR POLYAXIAL TULIP ASSEMBLIES (64-MT-0403) WERE THEN ATTACHED. A 40MM ROD (63-LT-5040) WAS PLACED IN THE TULIP. SET SCREWS (39-LS-0100) WERE THEN PLACED DOWN THE L4 TULIP UNTIL FORCE WAS INITIATED THAT WOULD RESULT IN THE FORCE LIMIT ON THE TORQUE LIMITING HANDLE TO RELEASE. UPON MAKING THE FINAL FORCED TURN, THE SHAFT OF THE DRIVER SHEARED OFF IN THE SET SCREW. THE DOCTOR WAS ABLE TO RETRIEVE THE BROKEN METAL. THE HANDLE/DRIVER COMBO WAS THEN EXCHANGED FROM A BACK-UP TRAY AND ALL FOUR (4) SET SCREW FUNCTIONED CORRECTLY. THERE WAS NO DELAY TO THE PROCEDURE RESULTING FROM THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18563362 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page