Reform Ti CT Modular MIS Pedicle Screw System 64-RD-9010Recalls & reported problems
Precision Spine, Inc. · Rod Inserter · Class 1
All models of Reform Ti CT Modular MIS Pedicle Screw System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840019942515), received between Dec 12, 2022 and Mar 1, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Break1
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMar 1, 2024Fracture; Fracture
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED IN AUSTRALIA ON (B)(6) 2023, UTILIZING THE REFORM TI MIS CT PEDICLE SCREW SYSTEM. DURING INSERTION OF A ROD, THE TIP OF THE ROD INSERTER ASSEMBLY (B)(6) BROKE WHILE TRYING TO MANIPULATE IT INTO POSITION. THE BROKEN FRAGMENT PIECE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED AS USUAL AS THE ROD WAS IN PLACE WHEN THE TIP BROKE. THERE WAS A 5 MINUTE DELAY TO THE PROCEDURE RESULTING FROM THE BROKEN INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18818735 ↗InjuryMar 1, 2024Fracture; Fracture
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED IN (B)(6) ON (B)(6), UTILIZING THE REFORM TI MIS CT PEDICLE SCREW SYSTEM. DURING INSERTION OF A ROD, THE TIP OF THE ROD INSERTER ASSEMBLY (64-RD-9010) BROKE WHILE TRYING TO MANIPULATE IT INTO POSITION. THE PROCEDURE WAS COMPLETED AS USUAL AS THE ROD WAS IN PLACE WHEN THE TIP BROKE. THERE WAS A 5-10 MINUTE DELAY TO THE PROCEDURE RESULTING FROM THE BROKEN INSTRUMENT. A SECOND SURGERY WAS PERFORMED ON AN UNKNOWN DATE TO RECOVER THE BROKEN FRAGMENT FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18817676 ↗InjuryDec 12, 2022Break; Entrapment of Device
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2022, UTILIZING THE REFORM MIS CT PEDICLE SCREW SYSTEM. WHEN TIGHTENING DOWN ON ONE OF THE LOCK SCREWS FOR THE SECOND ROD INSERTED, THE TIP OF THE ROD INSERTER ASSEMBLY REFORM MIS (39-RD-9010) BROKE. THE DOCTOR CONTINUED TIGHTENING DOWN ON THE ROD LEAVING THE BROKEN INSERTER TIP LOCKED UNDER THE ROD. THERE WAS NO DELAY TO THE PROCEDURE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15958933 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page