Reform Ti CT Modular MIS Pedicle Screw System 64-RD-9010Recalls & reported problems

Precision Spine, Inc. · Rod Inserter · Class 1

All models of Reform Ti CT Modular MIS Pedicle Screw System (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840019942515), received between Dec 12, 2022 and Mar 1, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Break1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED IN AUSTRALIA ON (B)(6) 2023, UTILIZING THE REFORM TI MIS CT PEDICLE SCREW SYSTEM. DURING INSERTION OF A ROD, THE TIP OF THE ROD INSERTER ASSEMBLY (B)(6) BROKE WHILE TRYING TO MANIPULATE IT INTO POSITION. THE BROKEN FRAGMENT PIECE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED AS USUAL AS THE ROD WAS IN PLACE WHEN THE TIP BROKE. THERE WAS A 5 MINUTE DELAY TO THE PROCEDURE RESULTING FROM THE BROKEN INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818735 ↗
  • InjuryMar 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED IN (B)(6) ON (B)(6), UTILIZING THE REFORM TI MIS CT PEDICLE SCREW SYSTEM. DURING INSERTION OF A ROD, THE TIP OF THE ROD INSERTER ASSEMBLY (64-RD-9010) BROKE WHILE TRYING TO MANIPULATE IT INTO POSITION. THE PROCEDURE WAS COMPLETED AS USUAL AS THE ROD WAS IN PLACE WHEN THE TIP BROKE. THERE WAS A 5-10 MINUTE DELAY TO THE PROCEDURE RESULTING FROM THE BROKEN INSTRUMENT. A SECOND SURGERY WAS PERFORMED ON AN UNKNOWN DATE TO RECOVER THE BROKEN FRAGMENT FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18817676 ↗
  • InjuryDec 12, 2022Break; Entrapment of Device

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2022, UTILIZING THE REFORM MIS CT PEDICLE SCREW SYSTEM. WHEN TIGHTENING DOWN ON ONE OF THE LOCK SCREWS FOR THE SECOND ROD INSERTED, THE TIP OF THE ROD INSERTER ASSEMBLY REFORM MIS (39-RD-9010) BROKE. THE DOCTOR CONTINUED TIGHTENING DOWN ON THE ROD LEAVING THE BROKEN INSERTER TIP LOCKED UNDER THE ROD. THERE WAS NO DELAY TO THE PROCEDURE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 15958933 ↗

Other Orthopedic Manual Surgical Instrument devices