ShurFit Lumbar Interbody 72-IN-1001Recalls & reported problems

Precision Spine, Inc. · Curved Cage Implant Inserter · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019960403), received between Jan 13, 2025 and Jan 13, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2025Break; Break

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2024, UTILIZING THE SHURFIT 3D TITANIUM INTERBODY CAGE SYSTEM. THE SURGEON WAS USING A MALLET TO PLACE THE CAGE INTO THE PROPER POSITION WHEN THE TIP OF TWO DIFFERENT ASSEMBLY, INSERTER, POSTERIOR CURVED CAGE (72-IN-1001) BROKE. THE BROKEN PIECES WITH A PICKUP AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY OPENING ANOTHER SET THAT WAS READILY AVAILABLE. THERE WAS NO PATIENT INJURY OR SURGICAL DELAY. IT WAS NOTED THAT THE SURGEON INDICATED THE PATIENT HAD VERY HARD BONE AND DOES NOT ALLEGE THE DEVICE HAD ANYTIHNG TO DO WITH THE INSTRUMENT FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21139086 ↗
  • MalfunctionJan 13, 2025Break; Break

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2024, UTILIZING THE SHURFIT 3D TITANIUM INTERBODY CAGE SYSTEM. THE SURGEON WAS USING A MALLET TO PLACE THE CAGE INTO THE PROPER POSITION WHEN THE TIP OF TWO DIFFERENT ASSEMBLY, INSERTER, POSTERIOR CURVED CAGE (72-IN-1001) BROKE. THE BROKEN PIECES WITH A PICKUP AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY OPENING ANOTHER SET THAT WAS READILY AVAILABLE. THERE WAS NO PATIENT INJURY OR SURGICAL DELAY. IT WAS NOTED THAT THE SURGEON INDICATED THE PATIENT HAD VERY HARD BONE AND DOES NOT ALLEGE THE DEVICE HAD ANYTHING TO DO WITH THE INSTRUMENT FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21139084 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →