Skip to content

Viant Medical, LLC T17652Recalls & reported problems

Viant Medical · STRAIGHT REAMER HANDLE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00840096400076), received between May 19, 2023 and Jul 2, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement11
  • Unstable11
  • Mechanical Problem3
  • Material Deformation2
  • Material Integrity Problem2
  • Material Twisted/Bent2

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2026Mechanical Problem; Unstable; Material Deformation; Material Integrity Problem; Material Twisted/Bent; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE REAMER HANDLE IS NO LONGER SPINNING CIRCULARLY. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25696593 ↗
  • MalfunctionMar 3, 2026Mechanical Problem; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE SURGEON NOTICED THE REAMER SHAFT WAS WOBBLY, POSSIBLY BENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24501428 ↗
  • MalfunctionFeb 12, 2026Mechanical Problem; Unstable; Unintended Movement

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT ON AN UNKNOWN PATIENT THAT THE REAMER SHAFT WAS WOBBLING WHEN CONNECTED TO THE STRYKER SYSTEM 9 ROTARY POWER. THE DEVICE WAS NOT USED BUT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24323013 ↗
  • MalfunctionJan 22, 2026Unstable; Unintended Movement

    IT WAS REPORTED DURING A TOTAL HIP PROCEDURE ON AN UNKNOWN PATIENT THAT THE STRAIGHT REAMER SHAFT IS DAMAGED AND NOT ROTATING STRAIGHT. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24145373 ↗
  • MalfunctionOct 29, 2025Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE CONNECTION POINT HAS WORN DOWN AND THE SHAFT IS WOBBLING WITH USE. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23415956 ↗
  • MalfunctionOct 29, 2025Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE CONNECTION POINT HAS WORN DOWN AND THE SHAFT IS WOBBLING WITH USE. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23415953 ↗
  • MalfunctionSep 19, 2025Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON A FEMALE PATIENT THAT THE REAMER IS SPINNING ELLIPTICALLY. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23096942 ↗
  • MalfunctionFeb 6, 2025Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE REAMER IS PULLING OUT OF THE CHUCK. THE FITTING IS WORN AND NO LONGER COUPLED WELL IN THE CHUCK. IT WOBBLES AS THE CONNECTION IS STRIPPING. IT WAS NOT USED FOR THE PROCEDURE AS ANOTHER DEVICE WAS AVAILABLE. THERE WAS A 30 SECOND SURGICAL DELAY, BUT NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21326791 ↗
  • MalfunctionMar 29, 2024Unstable; Material Deformation; Material Twisted/Bent; Unintended Movement; Unstable; Material Deformation; Material Twisted/Bent; Unintended Movement

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON A 70 YEAR OLD FEMALE PATIENT THAT THE REAMER SHAFT IS SLIGHTLY BENT, LEADING TO WOBBLE WHILE REAMING. NO PIECES BROKE OFF AND IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19004640 ↗
  • MalfunctionJan 4, 2024Unstable; Material Integrity Problem; Unintended Movement; Unstable; Material Integrity Problem; Unintended Movement

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT ON A FEMALE PATIENT THAT THE REAMER HANDLE WAS DAMAGED AT END OF HANDLE THAT CONNECTS TO POWER. THE REAMER WOBBLES BECAUSE OF IT. THERE WERE NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18445375 ↗
  • MalfunctionMay 19, 2023Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED DURING AN ARTHROPLASTY ON AN UNKNOWN PATIENT THAT WHILE USING THE QUICKSET ACE GRATER HANDLE IT WAS NOTED THAT THE GRATER ITSELF WAS WOBBLY AND NOT SITTING ON THE HANDLE CORRECTLY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE EVENTS NOR PATIENT CONSEQUENCES AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16961265 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates