Viant Medical, LLC T17653Recalls & reported problems
Viant Medical · OFFSET REAMER HANDLE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00840096400083), received between Apr 12, 2022 and Aug 8, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement8
- Break4
- Unstable4
- Vibration4
- Fracture3
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2025Break; Fracture; Material Integrity Problem; Break; Fracture; Material Integrity Problem
IT WAS REPORTED DURING A THA SURGERY FOR OA ON THE RIGHT HIP JOINT ON AN UNKNOWN PATIENT THAT THE PROCEDURE WAS TEMPORARILY HALTED AFTER DAMAGE TO THE SHAFT (INSIDE) WAS CONFIRMED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE WITH A 30 MINUTE SURGICAL DELAY. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22749870 ↗MalfunctionMar 12, 2025Break; Unstable; Material Integrity Problem; Unintended Movement; Break; Unstable; Material Integrity Problem; Unintended Movement
IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT WHILE REAMING ACETABULUM THE SURGEON NOTICED THAT THE REAMER WAS NOT REAMING SYMMETRICALLY. THE REAMER HANDLE WAS TAKEN APART AND THE INNER SHAFT OF THE REAMER WAS BROKEN. ALL BROKEN PIECES WERE ACCOUNTED FOR, REMOVED FROM THE FIELD, AND DISPOSED OF. THERE WAS 5 MINUTE SURGICAL DELAY IN SURGERY WHILE STAFF GOT ANOTHER TRAY. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21585594 ↗MalfunctionSep 24, 2024Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING HIP SURGERY THE REAMER BROKE.
Patient problems: Insufficient Information
MAUDE report 20291074 ↗MalfunctionJun 17, 2024Vibration; Material Deformation; Material Twisted/Bent; Unintended Movement; Vibration; Material Deformation; Material Twisted/Bent; Unintended Movement
IT WAS REPORTED DURING AN UNKNOWN PROCEDURE THAT THE REAMER HANDLE SHAFT WAS BENT AND DID NOT ALLOW REAMER TO SPIN ON THE SAME PLANE CONTINUOUSLY. NO KNOWN CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: Insufficient Information
MAUDE report 19547608 ↗MalfunctionJun 7, 2024Fracture; Vibration; Unintended Movement; Fracture; Vibration; Unintended Movement
IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE INSTRUMENT WOBBLED AND THEN FRACTURED INTO TWO PIECES. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19492002 ↗MalfunctionJun 7, 2024Fracture; Use of Device Problem; Vibration; Unintended Movement; Fracture; Use of Device Problem; Vibration; Unintended Movement
IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE INSTRUMENT WOBBLED AND THEN FRACTURED INTO TWO PIECES. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19484332 ↗MalfunctionMay 15, 2024Vibration; Unintended Movement; Vibration; Unintended Movement
IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE DEVICE IS WOBBLING IN THE DRILL CHUCK. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19320402 ↗MalfunctionOct 24, 2023Unstable; Unintended Movement; Unstable; Unintended Movement
IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT ON AN UNKNOWN PATIENT THAT THE OFFSET GRATER HANDLE HAD A WOBBLE WHILE REAMING. AFTER THE CASE THE INSTRUMENT WAS DISASSEMBLED AND INSPECTED AND FOUND THE UNIVERSAL JOINT HAD SIGNIFICANT WEAR AND WAS READY TO BREAK. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17995619 ↗MalfunctionSep 29, 2023Unstable; Unintended Movement; Unstable; Unintended Movement
IT WAS REPORTED DURING AN ANTERIOR TOTAL HIP REPLACEMENT ON AN UNKNOWN PATIENT THAT THE REAMER HANDLE APPEARED TO BE WOBBLY WHEN RUNNING ON POWER. AFTER BREAKING DOWN THE DEVICE, IT WAS NOTICED THE JOINTS ON THE INSIDE WERE WORN OUT. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17844167 ↗MalfunctionMay 5, 2023Unstable; Unintended Movement; Unstable; Unintended Movement
IT WAS REPORTED DURING ACETABULAR PREPARATION FOR A DIRECT ANTERIOR APPROACH (DAA) TOTAL HIP REPLACEMENT (THR) ON A FEMALE PATIENT THAT THE SURGEON NOTICED THERE WAS A LOT OF EXTRA MOVEMENT/PLAY IN THE DISTAL END OF THE OFFSET REAMER HANDLE. THE SURGEON CONTINUED TO USE THE HANDLE ON A LOW SPEED TO NOT AFFECT THE BONE-IMPLANT INTERFACE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE EVENTS NOR PATIENT CONSEQUENCE AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16877960 ↗MalfunctionApr 12, 2022Break; Break
IT WAS REPORTED THAT THE BLUE HANDLE OFFSET GRATEBATCH REAMER WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14084360 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.