FlowerToe Cannulated, D:2.5mm x 15mm FHT 250Recalls & reported problems
Flower Orthopedics Corporation · implant for orthopedic bone fixation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840118114066), received between Mar 9, 2021 and Mar 9, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2021Fracture; Fracture
"IN A PHONE CALL ON (B)(6) 2021, (B)(6) NOTED THAT TWO IMPLANTS BROKE DURING DR. (B)(6) CASE. THE FIRST BROKEN IMPLANT WAS DESCRIBED AS SNAPPING WHEN IT WAS BEING SCREWED IN, SPECIFICALLY AT THE HEAD OF THE IMPLANT WHERE "IT'S CONNECTED TO WHERE IT'S BROACHED." THIS IMPLANT WAS DISPOSED OF AND A NEW IMPLANT WAS EFFECTIVELY USED. THE SECOND IMPLANT ALSO BROKE IS A SIMILAR LOCATION WHEN TRYING TO SCREW IN THE IMPLANT. (B)(6) ALSO NOTED THAT THE IMPLANT TIP APPEARED BROKEN. HOWEVER, THE METAL THREADING WAS DESCRIBED AS ALSO "UNRAVELING" AND "DISINTEGRATING." THIS METAL THREADING CAME OFF OF THE IMPLANT AND WAS LODGED IN THE PATIENT. THE DOCTOR MADE AN INCISION NEAR THE IMPLANT SITE AND WAS ABLE TO SUCCESSFULLY REMOVE THE METAL THREADING WITHOUT USING X-RAY. THE DOCTOR WAS FRUSTRATED DUE TO THE TWO EVENTS AND ENDED UP USING K-WIRES FOR FIXATION AS AN ALTERNATIVE. IT WAS NOTED THAT THE SURGERY TIME INCREASED FROM 1.5-2 H TO 3.5-4 H DUE TO THE TWO EVENTS. HOWEVER, THE PATIENT WAS NOTED TO HAVE NO ADVERSE EVENTS AS OF THEIR FOLLOW-UP VISIT ON (B)(6) 2021. DR. (B)(6) DESCRIBED THE IMPLANTS AS BEING VERY "DELICATE" AND RECOMMENDED CREATING A MORE FLEXIBLE IMPLANT DESIGN."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11444790 ↗
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page