Skip to content

Nex-D2 CBZB005-45Recalls & reported problems

Next Orthosurgical, Inc. · BONE TAP, MIS, 4.5MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840314121103), received between Aug 9, 2025 and Jan 30, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device2
  • Material Fragmentation2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2026Break

    DURING LUMBAR SPINE SURGERY, A BONE TAP MIS 4.5 MM FRACTURED DURING NORMAL USE. THE DISTAL TIP OF THE BONE TAP REMAINED RETAINED WITHIN THE PATIENT'S BONE AND COULD NOT BE RETRIEVED INTRAOPERATIVELY. NO PATIENT INJURY, ADVERSE CLINICAL OUTCOME, OR ADDITIONAL MEDICAL OR SURGICAL INTERVENTION RELATED TO HARM WAS REPORTED. THE DEVICE DID NOT PERFORM AS INTENDED. THE EVENT IS CONSIDERED A DEVICE MALFUNCTION WITH POTENTIAL FOR SERIOUS INJURY IF IT WERE TO RECUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24222175 ↗
  • MalfunctionSep 8, 2025Material Fragmentation; Device Damaged by Another Device; Material Fragmentation; Device Damaged by Another Device

    DURING FIXATION OF L4 AND L5 VERTABREA THE TIP OF THE BONE TAP 4.5MM PN CBZB005-45 LN 0001217C1 BROKE WHILE TAPPING L5. A SNAPPING SOUND WAS HEARD AND FRAGMENTS OF THE TIP OF THE BONE TAP BECAME LODGED INSIDE THE BONE AND COULD NOT BE REMOVED. L5 WAS SKIPPED AND INSTEAD S1 WAS USED IN ITS PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22997742 ↗
  • MalfunctionAug 9, 2025Material Fragmentation; Device Damaged by Another Device; Material Fragmentation; Device Damaged by Another Device

    ON (B)(6) 2025, DURING A SURGICAL OPERATION PROCEDURE INVOLVING SPINAL FIXATION OF THE VERTEBRA T8 AND T9, COMPLICATIONS AROSE WITH THE INSTRUMENTATION. WHILE TAPPING THE VERTEBRAE, THE TIP OF THE INITIAL 4.5MM BONE TAP FRACTURED AND BECAME LODGED WITHIN THE T8 VERTEBRAL BODY. SUBSEQUENTLY, A SECOND 4.5MM BONE TAP WAS USED FOR T9, THE TIP OF THIS BONE TAP ALSO BECAME LODGED. RESULTING IN TWO SPERATE BONE TAP FRAGMENTS RETAINED WITHIN THE RESPECTIVE VERTEBRAE (T8 AND T9).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22766514 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page