INFINITY 33650015Recalls & reported problems
Wright Medical Technology, Inc. · Class 2 · Implant
All models of INFINITY (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00840420122568), received between Dec 30, 2021 and Apr 28, 2026 — classified as 14 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Osseointegration Problem3
- Loss of Osseointegration2
- Migration2
- Implant subsidence1
- Insufficient Device Problem Information1
Patient problems
- Inadequate Osseointegration5
- Unspecified Infection5
- Implant Pain4
- Insufficient Information2
- Osteolysis2
- Cyst(s)1
- Extraskeletal ossification1
- Osteopenia/ Osteoporosis1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryApr 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES FROM THE MONTHLY REPORT OF FEBRUARY 2026 DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25015839 ↗InjuryDec 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN (B)(6) 2025 TO (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION ACTIVE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 23687347 ↗InjuryJul 29, 2025Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Implant subsidence; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Implant subsidence
THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO EXOPHYTIC OSTEOLYTIC LESION ON THE ANTERIOR TIBIA.
Patient problems: Cyst(s); Extraskeletal ossification; Inadequate Osseointegration
MAUDE report 22656874 ↗InjuryApr 1, 2025Loss of Osseointegration; Loss of Osseointegration
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO COMPONENT MIGRATION/DISSOCIATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Inadequate Osseointegration
MAUDE report 21741695 ↗InjuryFeb 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL JANUARY 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO UNEXPLAINED PAIN, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Implant Pain
MAUDE report 21410808 ↗InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO UNEXPECTED PAIN, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: CONVERSION TO ARTHRODESIS.
Patient problems: Implant Pain
MAUDE report 21381892 ↗InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN JANUARY 2018 ¿ DECEMBER 2023. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE TO HIGH SUSPICION (E.G. PUS OR CONFIRMED MICRO) & LYSIS OF TALUS COMPONENT, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 20640315 ↗InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN (B)(6) 2018 ¿ (B)(6) 2023. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE TO HIGH SUSPICION (E.G. PUS OR CONFIRMED MICRO), WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 20638603 ↗InjuryNov 7, 2024Osseointegration Problem; Osseointegration Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN JANUARY 2018 ¿ DECEMBER 2023. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2018 A PATIENT REQUIRED REVISION SURGERY DUE TO HIGH SUSPICION (E.G. PUS OR CONFIRMED MICRO) & ASEPTIC LOOSENING OF TALAR COMPONENT, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 20634775 ↗InjuryOct 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "STUDY UK INFINITY® TOTAL ANKLE SYSTEM SUBJECT: (B)(6). PATIENT BLEEDING FROM LEFT ANKLE WOUND. PATIENT BECAME HYPOTENSIVE. 64/47- 1000MLS CRYSTALLOID ADMINISTERED AND BLOOD PRESSURE IMPROVED TO 78/59. PATIENT ALSO POSITIONED HEAD DOWN. PLASTER CUT AND PRESSURE AND COLD COMPRESS APPLIED TO BLEEDING POINT. BLEEDING STOPPED AND BLOOD PRESSURE IMPROVED WITH BLOOD TRANSFUSION AND CRYSTALLOID REPLACEMENT."
Patient problems: Wound Dehiscence
MAUDE report 20543321 ↗InjuryJul 18, 2024Osseointegration Problem; Migration; Osseointegration Problem; Migration
IT WAS REPORTED: PATIENT REQUIRES REVISION SURGERY DUE TO PAIN AND LOOSENING OF THE TIBIA AND TALUS COMPONENTS. LIKELY DUE TO INCREASED STRESS WITH SUBTALAR FUSION AND SCLEROTIC DISTAL TIBIA FROM PRIOR FRACTURE AND SURGERY. THE PLAN IS FOR FULL REVISION OF ALL IMPLANTS.
Patient problems: Implant Pain; Inadequate Osseointegration
MAUDE report 19777530 ↗InjuryNov 16, 2022Loss of Osseointegration; Insufficient Device Problem Information; Migration; Loss of Osseointegration; Insufficient Device Problem Information; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL ANKLE REPLACEMENT. ALLEGEDLY, THE PATIENT MAY NEED TO UNDERGO A REVISION SURGERY FOR REASONS THAT ARE NOT AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Inadequate Osseointegration; Insufficient Information; Osteopenia/ Osteoporosis
MAUDE report 15807678 ↗InjuryAug 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL ANKLE REPLACEMENT. ALLEGEDLY, THE PATIENT MAY NEED TO UNDERGO A REVISION SURGERY FOR REASONS THAT WERE NOT AVAILABLE AT THIS TIME.
Patient problems: Insufficient Information; Osteolysis
MAUDE report 15142486 ↗InjuryDec 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT CAME IN TO CLINIC AFTER 3 YEARS POST OP, COMPLAINING OF PAIN. THE PATIENT UNDERWENT A REVISION SURGERY.
Patient problems: Implant Pain
MAUDE report 13119500 ↗
Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Vantage 350-22-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Vantage 350-21-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls0 reports
- INFINITY 33630023Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER60 reports
- INFINITY 33600020Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Vantage 350-22-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall5 reports
- INFINITY 33650003Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER55 reports
- INFINITY 33630022Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER54 reports
- INFINITY 33600030Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page