Skip to content

INBONE 220222905Recalls & reported problems

Wright Medical Technology, Inc. · Class 2 · Implant

All models of INBONE (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840420124296), received between Feb 26, 2024 and Dec 30, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Osseointegration Problem2
  • Loss of Osseointegration1

Patient problems

  • Cyst(s)2
  • Inadequate Osseointegration2
  • Implant Pain1
  • Osteolysis1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ANKLE JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN (B)(6) 2018 ¿ (B)(6) 2023. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION LYSIS OF TALUS, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 21035786 ↗
  • InjuryFeb 28, 2024Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO INDICATION OF CHRONICALLY PAINFUL TOTAL ANKLE AND PRESUMED ASEPTIC LOOSENING. JOINT ASPIRATION CULTURES NEGATIVE. JOINT FLUID SYNOVASURE NEGATIVE. BASED ON EXAM, SERIAL RADIOGRAPHS OVER THE PAST 18 MONTHS AND SPECT-CT, REASONABLE SUSPICION FOR AT LEAST TALAR COMPONENT LOOSENING. THE PHYSICIAN INTENDS TO REVISE TALUS, AT A MINIMUM, BUT WILL ASSESS TIBIAL COMPONENT STABILITY INTRAOPERATIVELY.

    Patient problems: Cyst(s); Implant Pain; Inadequate Osseointegration

    MAUDE report 18800707 ↗
  • InjuryFeb 26, 2024Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY BECAUSE THE TIBIAL COMPONENT SINCE IT APPEARS LOOSE AND HAS MEDIAL MALLEOLAR CYST FORMATION. THE PHYSICIAN BROUGHT THE PATIENT TO SURGERY TO POSSIBLY REMOVE THE INBONE STEM AND TIBIAL TRAY AND TO REMOVE THE TALAR DOME. THE TIBIAL STEM AND TRAY WERE SOLID. THE PHYSICIAN USED THE INVISION DOME AND PLATE.

    Patient problems: Cyst(s); Inadequate Osseointegration

    MAUDE report 18783120 ↗

Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →