Skip to content

INBONE 220224310ERecalls & reported problems

Wright Medical Technology, Inc. · Class 2 · Implant

All models of INBONE (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840420125125), received between Nov 22, 2022 and Apr 26, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Osseointegration Problem1

Patient problems

  • Cyst(s)1
  • Inadequate Osseointegration1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 26, 2024Osseointegration Problem; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO TIBIAL COMPONENT LOOSENING AND POLYETHYLENE DEBRIS. THE LEFT ANKLE WAS ASPIRATED TODAY THROUGH A ANTEROMEDIAL APPROACH AND FLUID WAS SENT FOR TOTAL CELL COUNT, GRAM STAIN, CULTURE AND SENSITIVITY.

    Patient problems: Cyst(s); Inadequate Osseointegration

    MAUDE report 19195936 ↗
  • InjuryNov 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PHYSICIAN BELIEVED THAT HE DIDN¿T ENGAGE THE TIBIAL STEM TO THE TIBIAL TRAY CORRECTLY. THE TIBIAL TRAY CAME AWAY FROM THE STEM. IT WAS NOTED THAT INFECTION OCCURRED AROUND THE STEM. THE SURGEON EXPLANTED ALL THE COMPONENTS. THE PATIENT IS HEALING IN PREPARATION FOR FUSION.

    Patient problems: Unspecified Infection

    MAUDE report 15843453 ↗

Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →