Skip to content

Valor® 415101120LRecalls & reported problems

Wright Medical Technology, Inc. · Class 2 · Implant

All models of VALOR (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840420125583), received between Nov 7, 2021 and Feb 14, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Remove1
  • Material Deformation1
  • Migration1

Patient problems

  • Inflammation1
  • Insufficient Information1
  • Nonunion/Delayed-union Bone Fracture1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 14, 2022Migration; Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED REDNESS AND SWELLING LOWER LEG AND ANKLE. IT WAS ALSO NOTED THAT THE IM NAIL LOOSENED AND CAUSED INFECTION THAT LED TO A NONUNION OF THE ANKLE FUSION SITE. IT REQUIRED REMOVAL OF THE IM NAIL WITH BONE BIOPSIES AND PLACEMENT OF ABX CEMENT NAIL. IN ADDITION TO THAT THE PATIENT WAS GIVEN IV ANTIBIOTICS FOR 6 WEEKS FOLLOWED WITH REVISION OF ANKLE FUSION WITH EXTERNAL FIXATOR.

    Patient problems: Inflammation; Nonunion/Delayed-union Bone Fracture; Unspecified Infection

    MAUDE report 13524844 ↗
  • InjuryNov 7, 2021Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGERY TO REMOVE THE NAIL. ALLEGEDLY, THE COMPRESSION SCREW IN THE NAIL WAS DEFECTIVE WHICH MADE IT IMPOSSIBLE TO SCREW THE EXTRACTOR ONTO THE NAIL TO EXTRACT THE NAIL CORRECTLY. THE SURGEON MADE TWO ADDITIONAL INCISIONS AT THE TIBIA THAT WERE NOT PLANNED TO REMOVE THE NAIL. THIS RESULTED IN THE SURGERY LASTING FOR 3 HOURS INSTEAD OF 45 MINUTES. IT ALSO RESULTED IN 3 MONTHS OF CONVALESCENCE INSTEAD OF 2 WEEKS.

    Patient problems: Insufficient Information

    MAUDE report 12763301 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page