ORTHOLOC 3Di 587111RTRecalls & reported problems

Wright Medical Technology, Inc. · Class 2

All models of ORTHOLOC 3Di (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840420126511), received between Jun 5, 2023 and Sep 5, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1
  • Positioning Failure1

Patient problems

  • Granuloma1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2024Positioning Failure; Positioning Failure

    AS REPORTED BY SALES AGENT: "SURGEON PUT PLATE ON AND FIXED WITH ONE SCREW DISTALLY. WENT TO PUT SECOND LOCKING SCREW IN AND SAID SCREW WOULDN¿T LOCK IN. TRIED SEVERAL METHODS TO LOCK IN BUT SURGEON BELIEVES THREADS ON PLATES FAULTY. REMOVED PLATE AND SWAPPED TO STANDARD ORTHOLOC MTP FUSION PLATE. NO FURTHER ISSUES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20153766 ↗
  • InjuryJun 5, 2023Material Fragmentation; Material Fragmentation

    THE ORTHOLOC SYSTEM, COMPOSED OF PLATES AND SCREWS, USED FOR ARTHRODESIS OF THE FIRST RAY OF THE FOOT FOR HALLUX RIGIDUS, RELEASED A METAL FRAGMENT WHICH CAUSED A FOREIGN BODY GRANULOMA IN THE SOFT TISSUE. THIS WAS NOTICED POST-OP ON X-RAY DUE TO PRESENCE OF PAIN. SURGERY WAS NECESSARY TO REMOVE THE GRANULOMA, THE METAL FRAGMENT WAS IDENTIFIED INSIDE IT.

    Patient problems: Granuloma; Pain

    MAUDE report 17058626 ↗