enableCV IntraClude™ Intra-Aortic Occlusion Device ICF100Recalls & reported problems
Enablecv Inc. · IntraClude™ Intra-Aortic Occlusion Device - ICF100 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840479401010), received between Jun 12, 2026 and Jun 12, 2026 — classified as 1 death.
The most frequently reported specific device problem is “Use of Device Problem” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K132175, K163693); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem1
Patient problems
- Aortic dissection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathJun 12, 2026Use of Device Problem
A EVENT LEADING TO DEATH WAS REPORTED TO BE POSSIBLY ASSOCIATED WITH THE USE OF THE ER21B CANNULA AND ICF100 CATHETER. THE PHYSICIAN INITIALLY THOUGHT THE PATIENT'S VESSEL WAS TOO SMALL, BUT THEN THE SURGEON CUT THE VESSEL DOWN AND DECIDED TO MOVE FORWARD WITH THE USE OF THE ICF100. IT WAS REPORTED THAT THE GUIDEWIRE POSSIBLY GOT CAUGHT AT SOME POINT IN THE VESSEL (IT IS UNCLEAR IF THAT IS THE WIRE FOR THE ENDORETURN OR THE ICF100). THE ICF100 WAS NOT YET INFLATED. IT ALSO IS PRESUMED THAT SURGICAL TECHNIQUE OR AN ANOMALY WITH THE WIRE COULD HAVE BEEN RELATED TO THE EVENT, THOUGH NO ABNORMALITIES OR ANOMALLIES WERE REPORTED. NO RESISTANCE MENTIONED DURING ADVANCEMENT. ISSUE WAS NOT NOTICED UNTIL THE ICF100 WAS IN THE ASCENDING (PRIOR TO INFLATION). PREOP SCAN WAS WHEN THE PATIENT WAS SITTING UP SO THE DOCTOR DIDN'T GET A GOOD FEMORAL VIEW PRIOR TO THE PROCEDURE. ULTRASOUND BEFORE PROCEDURE OF FEMORAL VESSELS WAS APPROXIMATE FOR MEASUREMENT (6.8 MM)- WHEN THEY DID THE CUT DOWN, THEY THOUGHT THE VESSEL WAS LARGE ENOUGH. SURGEON CANNULATED WITH ER21B FOLLOWED BY PLACEMENT OF INTRACLUDE. SHORTLY AFTER LEFT AND RIGHT RADIAL PRESSURES WENT TO 0 AND TEE CONFIRMED A RETROGRADE AORTIC DISSE
Patient problems: Aortic dissection
MAUDE report 25486828 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132175, K163693. One submission often covers many models, so check the product and lot details in each record.
- Z-1705-2019 · May 14, 2019 · TerminatedIntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Sterile EO The IntraClude intra-aortic occlusion device is indicated for use
- Z-1100-2019 · Jan 16, 2019 · Open, ClassifiedEdwards Lifesciences(R) IntraClude(TM) Intra-Aortic Occlusion Device is a 10.5 Fr (3.5mm), triple lumen, 100cm long catheter with an elastomeric balloon near its distal tip des
- Z-1929-2018 · Mar 30, 2018 · TerminatedEdwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending aorta.
- Z-1758-2015 · May 5, 2015 · TerminatedEdwards Lifesciences IntraClude Intra-Aortic Occlusion Device, model number ICF100, 10.5 Fr (3.5 mm) x 39.4" (100 cm) ThruPort System. Packaged sterile and non-pyrogenic in a seale
Other Clamp, Vascular devices
- Heartstring III System 3.8MM HSK-3038Maquet Cardiovascular LLC · Clamp, Vascular1 recall4 deaths809 reports
- Heartstring III System 4.3MM HSK-3043Maquet Cardiovascular LLC · Clamp, Vascular1 recall265 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR284 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR256 reports
- HeartString III Proximal Seal HS-3045Maquet Cardiovascular LLC · Clamp, Vascular1 recall187 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR67 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR65 reports
- Flo-Thru Intraluminal Shunt® FT-12225Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page