Skip to content

LeMaitre Embolectomy Catheter e1651-38Recalls & reported problems

Lemaitre Vascular, Inc. · 3F 80CM Over-the-Wire Embolectomy Catheter, eIFU · Class 2

All models of LeMaitre Embolectomy Catheter (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840663100774), received between Dec 17, 2020 and Dec 17, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Material Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

10 other recalls cover devices cleared under the same FDA submission (K022145); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Apr 11, 2025Class IIOpen, Classified

    During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

    Z-1750-2025 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 17, 2020Material Separation; Material Separation

    TIP BROKE OFF INTO THE PATIENT. NO INJURY TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11029420 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K022145. One submission often covers many models, so check the product and lot details in each record.

  • Z-0117-2027 · Aug 28, 2026 · Open, Classified
    TufTex Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
  • Z-1567-2020 · Mar 23, 2020 · Open, Classified
    LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the r
  • Z-0378-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 3F, 40cm, Model 1651-34 and e1651-34. Indicated for the removal of arterial emboli and thrombi.
  • Z-0384-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 6F, 40 cm, Model e-1651-64. Indicated for the removal of arterial emboli and thrombi.
  • Z-0385-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 6F, 80 cm, Model e-1651-68. Indicated for the removal of arterial emboli and thrombi.
  • Z-0382-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 7F, 80cm, Model 1654-78 and e1654-78. Indicated for the removal of arterial emboli and thrombi.
  • Z-0381-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 80cm, Model 1651-88 and e1654-88. Indicated for the removal of arterial emboli and thrombi.
  • Z-0380-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 4F, 80cm, Model 1651-48 and e1654-48. Indicated for the removal of arterial emboli and thrombi.
  • Z-0383-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 40 cm, Model e1651-84. Indicated for the removal of arterial emboli and thrombi.
  • Z-0379-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 3F, 80cm, Model 1651-38. Indicated for the removal of arterial emboli and thrombi.

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page