LeMaitre Embolectomy Catheter 1801-38Recalls & reported problems
Lemaitre Vascular, Inc. · 3F 80 cm NovaSil Silicone Single Lumen Embolectomy Catheter · Class 2
All models of LeMaitre Embolectomy Catheter (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840663100903), received between Jun 3, 2019 and Jan 17, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
9 other recalls cover devices cleared under the same FDA submission (K992368); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED WHILE THE DEVICE WAS BEING PREPARED BY REMOVING THE AIR, THE BALLOON RUPTURED. ANOTHER DEVICE WAS OPENED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21179198 ↗MalfunctionOct 3, 2023Material Rupture; Material Rupture
IT WAS REPORTED THE SYNTEL SILICONE EMBOLECTOMY CATHETER BALLOON RUPTURED DURING A PROCEDURE. THE BALLOON WAS NOT OVER INFLATED, BUT RATHER UNDER INFLATED. THERE WERE NO CALCIUM BUILD UP IN THE PATIENT'S VESSEL. THE SURGEONS WERE CONVINCED IT WAS A PRODUCT MANUFACTURING ISSUE WITH NO ERROR IN USAGE OR PATIENT CONDITION. A 12TLW804F FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER MANUFACTURED BY BAXTER WAS USED TO COMPLETE THE PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT. NOTE: THIS IS REPORT 9 OF 9 RELATED TO THE NINTH CATHETER USED IN THE EVENT. REPORT 1220948-2023-00160, 1220948-2023-00161, 1220948-2023-00162, 1220948-2023-00163, 1220948-2023-00164, 1220948-2023-00165, 1220948-2023-00166, 1220948-2023-00167 WAS SUBMITTED FOR THE FIRST, SECOND, THIRD, FOURTH, FIFTH, SIXTH, SEVENTH, AND EIGHTH DEVICES INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17862300 ↗MalfunctionOct 3, 2023Material Rupture; Material Rupture
IT WAS REPORTED THE SYNTEL SILICONE EMBOLECTOMY CATHETER BALLOON RUPTURED DURING A PROCEDURE. THE BALLOON WAS NOT OVER INFLATED, BUT RATHER UNDER INFLATED. THERE WERE NO CALCIUM BUILD UP IN THE PATIENT'S VESSEL. A 12TLW804F FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER MANUFACTURED BY BAXTER WAS USED TO COMPLETE THE PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT. NOTE: THIS IS REPORT 1 OF 9 RELATED TO THE FIRST CATHETER USED IN THE EVENT. REPORT 1220948-2023-00161, 1220948-2023-00162, 1220948-2023-00163, 1220948-2023-00164, 1220948-2023-00165, 1220948-2023-00166, 1220948-2023-00167, 1220948-2023-00168 WAS SUBMITTED FOR THE SECOND, THIRD, FOURTH, FIFTH, SIXTH, SEVENTH, EIGHTH, AND NINTH DEVICES INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17862258 ↗MalfunctionJun 3, 2019Material Rupture; Material Rupture
THE SURGEON REPORTED THAT THE BALLOONS OF THE LEMAITRE NOVASIL SINGLE LUMEN EMBOLECTOMY CATHETERS BURST DURING EMBOLECTOMY PROCEDURE WHEN THE BALLOONS WERE EXPOSED TO HIGHLY CALCIFIED VESSEL. THIS IS REPORT 2 OF 3 OF THAT SERIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8664930 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K992368. One submission often covers many models, so check the product and lot details in each record.
- Z-0228-2023 · Sep 30, 2022 · Open, ClassifiedTufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by thi
- Z-0390-2016 · Oct 21, 2015 · TerminatedLeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 2F, 60 cm, Model e1801-26. Indicated for the removal of arterial emboli and thrombi.
- Z-0388-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 5F, 80 cm, Model 1601-58 and e1601-58. Indicated for the removal of arterial emboli and thrombi.
- Z-0389-2016 · Oct 21, 2015 · TerminatedLeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 6F, 80 cm, Model 1601-68. Indicated for the removal of arterial emboli and thrombi.
- Z-0386-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 3F, 80 cm, Model 1601-38 and e1601-38. Indicated for the removal of arterial emboli and thrombi.
- Z-0387-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 4F, 80 cm, Model 1601-48 and e1601-48. Indicated for the removal of arterial emboli and thrombi.
- Z-0558-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 3F Single Lumen, Balloon volume 0.2ml, Length 80 cm, Latex Model: 1601-38
- Z-0557-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 2F Single Lumen, Balloon Volume 0.05 ml, Length 60 cm. Latex Model: 1601-26
- Z-0560-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 5F Single Lumen, Balloon Volume 1.5 ml, Length 80 cm, Latex Model: 1601-58
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page