LeMaitre Embolectomy Catheter e1801-48Recalls & reported problems
Lemaitre Vascular, Inc. · 4F 80 cm NovaSil Silicone Single Lumen Embolectomy Catheter, eIFU · Class 2
All models of LeMaitre Embolectomy Catheter (47)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840663101047), received between Aug 9, 2019 and May 12, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
9 other recalls cover devices cleared under the same FDA submission (K992368); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Difficult to Remove1
- Failure to Deflate1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE SYNTEL SILICONE EMBOLECTOMY CATHETER BALLOON RUPTURED WHILE ATTEMPTING TO REMOVE A CLOT DURING A PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22001370 ↗MalfunctionAug 11, 2023Material Rupture; Material Rupture
IT WAS REPORTED SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP BALLOON RUPTURED IN VESSEL DURING BYPASS PROCEDURE. NO INJURY REPORTED TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17508195 ↗MalfunctionJul 7, 2023Difficult to Remove; Failure to Deflate; Difficult to Remove; Failure to Deflate
IT WAS REPORTED SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP BALLOON NOT DEFLATING AND STUCK IN PATIENT DURING DE-CLOT ACCESS CENTER PROCEDURE. THE BALLOON HAD TO BE PHYSICALLY POPPED WITH A NEEDLE. NO INJURY OCCURRED TO PATIENT. IT WAS ALSO STATED CUSTOMER HAS BEEN HAVING A LOT OF BALLOON RUPTURE DURING USE WITH PATIENT, BUT NO FURTHER INFORMATION (DATE AND LOT NUMBERS) WERE AVAILABLE. THE DEVICE WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17277731 ↗MalfunctionAug 9, 2019Material Rupture; Material Rupture
THE SURGEON REPORTED THAT DURING THROMBECTOMY ON ONE PATIENT, THE BALLOONS OF THE THREE LEMAITRE EMBOCTOMY CATHETERS RUPTURED. TWO WERE SILICONE BASED BALLOONS AND ONE WAS A LATEX BALLOON. THIS IS REPORT 2 OF 3 AND DESCRIBES A SILICONE BASED BALLOON FAILURE. REPORT 1 OF 3 DESCRIBES ANOTHER SILICONE BASED BALLOON FAILURE AND 3 OF 3 DESCRIBES A LATEX BASED BALLOON FAILURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8880775 ↗MalfunctionAug 9, 2019Material Rupture; Material Rupture
THE SURGEON REPORTED THAT DURING THROMBECTOMY ON ONE PATIENT, THE BALLOONS OF THE THREE LEMAITRE EMBOCTOMY CATHETERS RUPTURED. TWO WERE SILICONE BASED BALLOONS AND ONE WAS A LATEX BALLOON. THIS IS REPORT 1 OF 3 AND DESCRIBES A SILICONE BASED BALLOON FAILURE. REPORT 2 OF 3 DESCRIBES ANOTHER SILICONE BASED BALLOON FAILURE AND 3 OF 3 DESCRIBES A LATEX BASED BALLOON FAILURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8880772 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K992368. One submission often covers many models, so check the product and lot details in each record.
- Z-0228-2023 · Sep 30, 2022 · Open, ClassifiedTufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by thi
- Z-0387-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 4F, 80 cm, Model 1601-48 and e1601-48. Indicated for the removal of arterial emboli and thrombi.
- Z-0390-2016 · Oct 21, 2015 · TerminatedLeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 2F, 60 cm, Model e1801-26. Indicated for the removal of arterial emboli and thrombi.
- Z-0389-2016 · Oct 21, 2015 · TerminatedLeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 6F, 80 cm, Model 1601-68. Indicated for the removal of arterial emboli and thrombi.
- Z-0388-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 5F, 80 cm, Model 1601-58 and e1601-58. Indicated for the removal of arterial emboli and thrombi.
- Z-0386-2016 · Oct 21, 2015 · TerminatedLeMaitre Single Lumen Embolectomy Catheter 3F, 80 cm, Model 1601-38 and e1601-38. Indicated for the removal of arterial emboli and thrombi.
- Z-0558-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 3F Single Lumen, Balloon volume 0.2ml, Length 80 cm, Latex Model: 1601-38
- Z-0560-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 5F Single Lumen, Balloon Volume 1.5 ml, Length 80 cm, Latex Model: 1601-58
- Z-0557-05 · Feb 7, 2005 · TerminatedLeMaitre Embolectomy Catheter 2F Single Lumen, Balloon Volume 0.05 ml, Length 60 cm. Latex Model: 1601-26
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page