AlboGraft Polyester Vascular Graft AMC1608Recalls & reported problems
Lemaitre Vascular, Inc. · AlboGraft Polyester Vascular Graft 16cmx8mm · Class 2 · Implant
All models of AlboGraft Polyester Vascular Graft (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840663103454), received between Jun 17, 2016 and Jun 26, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Fluid/Blood Leak1
- Leak/Splash1
- Material Puncture/Hole1
Patient problems
- Blood Loss1
- Hemorrhage/Blood Loss/Bleeding1
- Irritation1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE ALBOGRAFT WALL WAS LEAKING AFTER BEING CLOSED WITH SUTURES FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. LEAKAGE OCCURRED FROM THE GRAFT RESULTING IN LONGER OPERATION TIME AND BLOOD LOSS. AN ALTERNATE GRAFT WAS USED TO COMPLETE THE PROCEDURE. NO FURTHER COMPLICATION OR INJURY WAS REPORTED TO THE PATIENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19617006 ↗InjuryJun 10, 2020Material Puncture/Hole; Material Puncture/Hole
INTRAOPERATIVELY, A HOLE WAS NOTICED IN THE GRAFT. THE HOLE WAS 3-4MM IN SIZE. ON ONE PICTURE IS A 1.2MM DULL- HEADED CANNULA IN THE HOLE OF THE GRAFT. THE HOLE WAS SEWN OVER. THE PATIENT WAS STILL IN THE OPERATION ROOM, WHEN THEY CALLED US. THERE WAS NO IMPACT ON PATIENT'S HEALTH AS THE RESULT OF THIS INCIDENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10138111 ↗InjuryDec 13, 2018Adverse Event Without Identified Device or Use Problem
A FEW YEARS AGO ( EXACT DATE UNKNOWN), A PATIENT RECEIVED A BIFURCATED AORTIC PROSTHESIS ( FROM A DIFFERENT MANUFACTURER ). THE PATIENT WAS DIAGNOSED WITH A PERIGRAFT SEROMA FROM THIS GRAFT. ON (B)(6) 2017, THE PROSTHESIS WAS REPLACED WITH AN ALBOGRAFT PROSTHESIS ( CATALOG# AMC1608; LOT# 180381 SERIAL NUMBER# (B)(4)). THE PATIENT HAD AN ALLERGIC REACTION TO THIS GRAFT AS WELL. SO, THE PATIENT WAS TREATED FOR AN ALLERGIC REACTION. THE PATIENT RESPONDED WELL TO MEDICATION. HOWEVER, SINCE THE BEGINNING OF (B)(6) 2018, THE PATIENT DID NOT RESPOND WELL TO THE PRESCRIBED MEDICATION. A BLOOD TEST CONDUCTED ON THE PATIENT'S BLOOD SAMPLE DETECTED PRESENCE OF METHY- PROPYL-4-HYDROXYBENZOATE.
Patient problems: Irritation
MAUDE report 8162158 ↗InjuryJun 17, 2016Leak/Splash; Leak/Splash
THE GRAFT WAS LEAKING DURING THE IMPLANTATION. THE PHYSICIAN HAD TO USE THE BLOOD CIRCULATING MACHINE FOR BLOOD TRANSFUSION (ABOUT 800 CC). THE BLEEDING WAS STOPPED WITH CELLULOSE WADDING. THE GRAFT REMAINS IMPLANTED. PATIENT IS OKAY.
Patient problems: Blood Loss
MAUDE report 5732537 ↗
Other Graft, Vascular, Synthetic/Biologic Composite devices
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
- Hemashield M002020851890Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall2 reports
- Hemashield M002020952070Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall0 reports
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- AlboGraft Polyester Vascular Graft AMC1407Lemaitre Vascular, Inc. · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page