AlboGraft Polyester Vascular GraftRecalls & reported problems across all models
Lemaitre Vascular, Inc. · 11 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls
- 0
We found 11 models of AlboGraft Polyester Vascular Graft by Lemaitre Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 22 adverse event reports in the FDA's MAUDE database, received between May 5, 2016 and Jul 17, 2024 — classified as 10 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported AlboGraft Polyester Vascular Graft problems
Device problems
- Fluid/Blood Leak5
- Leak/Splash4
- Material Puncture/Hole4
- Patient Device Interaction Problem2
- Adverse Event Without Identified Device or Use Problem1
- Device Contaminated at the User Facility1
- Device Issue1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient4
- Aneurysm3
- Blood Loss3
- Insufficient Information2
- No Consequences Or Impact To Patient2
- Failure of Implant1
- Hemorrhage/Blood Loss/Bleeding1
AlboGraft Polyester Vascular Graft models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AlboGraft Polyester Vascular Graft AMC1407Graft, Vascular, Synthetic/Biologic Composite6 reports
- AlboGraft Polyester Vascular Graft AMC1608Graft, Vascular, Synthetic/Biologic Composite4 reports
- AlboGraft Polyester Vascular Graft AMC3006Graft, Vascular, Synthetic/Biologic Composite2 reports
- AlboGraft Polyester Vascular Graft AMC1207Graft, Vascular, Synthetic/Biologic Composite2 reports
- AlboGraft Polyester Vascular Graft AMC2010Graft, Vascular, Synthetic/Biologic Composite2 reports
- AlboGraft Polyester Vascular Graft AMC4008Graft, Vascular, Synthetic/Biologic Composite1 report
- AlboGraft Polyester Vascular Graft AMC1809Graft, Vascular, Synthetic/Biologic Composite1 report
- AlboGraft Polyester Vascular Graft ATC1520Graft, Vascular, Synthetic/Biologic Composite1 report
- AlboGraft Polyester Vascular Graft ATC3028Graft, Vascular, Synthetic/Biologic Composite1 report
- AlboGraft Polyester Vascular Graft ATC2211Graft, Vascular, Synthetic/Biologic Composite1 report
- AlboGraft Polyester Vascular Graft ATC3036Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 report
Recent AlboGraft Polyester Vascular Graft adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2024Material Disintegration; Material Disintegration
IT WAS REPORTED THAT THE MATERIAL OF THE GRAFT STARTED TO DISINTEGRATE AS THE SUTURE WENT THROUGH IT; FURTHER SUTURING MADE THE GRAFT DISINTEGRATE MORE. ANOTHER GRAFT WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY. NO INJURY WAS REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19761147 ↗InjuryJun 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE ALBOGRAFT WALL WAS LEAKING AFTER BEING CLOSED WITH SUTURES FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. LEAKAGE OCCURRED FROM THE GRAFT RESULTING IN LONGER OPERATION TIME AND BLOOD LOSS. AN ALTERNATE GRAFT WAS USED TO COMPLETE THE PROCEDURE. NO FURTHER COMPLICATION OR INJURY WAS REPORTED TO THE PATIENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19617006 ↗MalfunctionApr 18, 2023Device Contaminated at the User Facility; Device Contaminated at the User Facility
DURING A MEETING WITH THE SWEDISH MOH WE WERE INFORMED OF AN INCIDENT WHERE WHILE PREPARING THE ALBOGRAFT, THE NURSE OVERLOOKED THE LABELS ON THE PACKAGING AND INTRODUCED THE NON-STERILE PACKAGING INTO THE STERILE FIELD. THE STATUS OF THE PATIENT IS UNKNOWN. THE NURSE DID NOT REPORT THE EVENT AT THE TIME OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16762228 ↗MalfunctionSep 20, 2022Material Puncture/Hole; Material Puncture/Hole
AN IMPLANT OF ALBOGRAFT AMC3008 FOR BYPASS WAS IN PROCESS. AT THE DISTAL PART OF THE GRAFT, THEY FOUND A SMALL HOLE. THIS PART WAS NOT AIMED TO BE USED, SO IT WAS CUT OFF AND WAS NOT USED. CONCLUSION A SMALL HOLE IN THE GRAFT WAS RECOGNIZED, BUT IT WAS NOT IN CONTACT OR BEEN USED IN THE PATIENT. THE GRAFT WAS DISCARDED BY THE HOSPITAL. THERE WAS NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15456844 ↗MalfunctionSep 2, 2021Material Puncture/Hole; Material Puncture/Hole
DURING AN INFRARENAL AORTIC REPLACEMENT, AFTER COMPLETING PROXIMAL ANASTOMOSIS, THE SURGEON OBSERVED A DEFECT IN THE Y SECTION OF THE GRAFT. HE OBSERVED A HOLE IN BETWEEN THE TWO BIFURCATES. IT WAS ABOUT 1 MM WITH A VISIBLE MESH DEFECT. THE HOLE WAS CLOSED WITH A SUTURE. NO OTHER DEFECTS WERE PRESENT IN THE GRAFT. THERE WAS NO HARM TO THE USER OR THE PATIENT BECAUSE OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12416062 ↗InjuryApr 22, 2021Material Puncture/Hole; Material Puncture/Hole
DURING PROCEDURE, AFTER COMPLETING DISTAL ANASTOMOSIS, SURGEON OBSERVED A SMALL HOLE IN ONE OF THE BIFURCATED LEG WHEN FLUSHING IT WITH SALINE. THE SECTION OF THE GRAFT WITH THE HOLE WAS DISCARDED. THERE WAS NO HARM TO THE PATIENT AS THE RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11711245 ↗InjuryNov 5, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
PROSTHESIS THROMBOSIS. FEM-POP ARTERIAL BYPASS. NEED OF ANOTHER SURGERY TO REPLACE IT.
Patient problems: Insufficient Information
MAUDE report 10795871 ↗InjuryNov 5, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
PROSTHESIS THROMBOSIS. FEM-POP ARTERIAL BYPASS. NEED OF ANOTHER SURGERY TO REPLACE IT.
Patient problems: Insufficient Information
MAUDE report 10795852 ↗InjuryAug 14, 2020Material Rupture; Material Rupture
SPONTANEOUS VERTICAL RUPTURE OF THE GRAFT IN THE AREA OF THE PROXIMAL UPPER THIRD IN LENGTH, AWAY FROM THE ANASTOMOSIS.
Patient problems: Failure of Implant
MAUDE report 10408881 ↗InjuryJun 10, 2020Material Puncture/Hole; Material Puncture/Hole
INTRAOPERATIVELY, A HOLE WAS NOTICED IN THE GRAFT. THE HOLE WAS 3-4MM IN SIZE. ON ONE PICTURE IS A 1.2MM DULL- HEADED CANNULA IN THE HOLE OF THE GRAFT. THE HOLE WAS SEWN OVER. THE PATIENT WAS STILL IN THE OPERATION ROOM, WHEN THEY CALLED US. THERE WAS NO IMPACT ON PATIENT'S HEALTH AS THE RESULT OF THIS INCIDENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10138111 ↗InjuryDec 13, 2018Adverse Event Without Identified Device or Use Problem
A FEW YEARS AGO ( EXACT DATE UNKNOWN), A PATIENT RECEIVED A BIFURCATED AORTIC PROSTHESIS ( FROM A DIFFERENT MANUFACTURER ). THE PATIENT WAS DIAGNOSED WITH A PERIGRAFT SEROMA FROM THIS GRAFT. ON (B)(6) 2017, THE PROSTHESIS WAS REPLACED WITH AN ALBOGRAFT PROSTHESIS ( CATALOG# AMC1608; LOT# 180381 SERIAL NUMBER# (B)(4)). THE PATIENT HAD AN ALLERGIC REACTION TO THIS GRAFT AS WELL. SO, THE PATIENT WAS TREATED FOR AN ALLERGIC REACTION. THE PATIENT RESPONDED WELL TO MEDICATION. HOWEVER, SINCE THE BEGINNING OF (B)(6) 2018, THE PATIENT DID NOT RESPOND WELL TO THE PRESCRIBED MEDICATION. A BLOOD TEST CONDUCTED ON THE PATIENT'S BLOOD SAMPLE DETECTED PRESENCE OF METHY- PROPYL-4-HYDROXYBENZOATE.
Patient problems: Irritation
MAUDE report 8162158 ↗MalfunctionOct 10, 2018Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem
DURING CAROTID ENDARTERECTOMY, THE INTERNAL CAROTID BALLOON FAILED TO INFLATE PROPERLY. AS A RESULT, BLOOD WAS LEAKING OUT FROM THE PATIENT'S VESSEL DURING ANASTOMOSIS SINCE THE BALLOON WAS UNABLE TO COMPLETELY OCCLUDE THE ARTERY. THIS IS A FOLLOW-UP REPORT OF MANUFACTURER'S INCIDENT REPORT 1220948-2018-00059 THAT WAS INITIALLY PROVIDED ON (B)(6) 2018.
Patient problems: Aneurysm
MAUDE report 7952761 ↗MalfunctionMay 17, 2018Fluid/Blood Leak; Fluid/Blood Leak
BLOOD WAS OOZING FROM LEFT LIMB OF THE GRAFT NEAR THE SITE OF BIFURCATION ONCE THE SUPPORTING CLAMP WAS REMOVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7523406 ↗InjuryMay 16, 2018Fluid/Blood Leak; Fluid/Blood Leak
DURING ABDOMINAL ANEURYSM REPAIR, AFTER THE ANASTOMOSIS WAS PERFORMED DISTALLY AND THE AORTIC CLAMP WAS REMOVED, THE BLOOD LEAKED THROUGH THE ENTIRE LENGTH OF THE GRAFT.
Patient problems: Blood Loss
MAUDE report 7519388 ↗InjuryDec 15, 2017Fluid/Blood Leak; Fluid/Blood Leak
ALBOGRAFT VASCULAR GRAFT AMC2010 LOT# 140129 SERIAL NUMBER (B)(4) WAS USED DURING OPEN AORTA ANEURISM TREATMENT. AFTER PERFORMING PROXIMAL ANASTOMOSIS, THE GRAFT START TO BLEED WHEN THE CLAMP WAS REMOVED. SO, ANOTHER LEMAITRE ALBOGRAFT WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 7123279 ↗
Other Lemaitre Vascular, Inc. products
- LeMaitre Embolectomy Catheter47 models · 438 reports · 2 recalls
- Pruitt F3 Carotid Shunt10 models · 305 reports
- Artegraft Vascular Graft15 models · 25 reports · 3 recalls
- Trivex System3 models · 69 reports
- Syntel Silicone OTW Embolectomy Catheter8 models · 66 reports
- Syntel Silicone Thrombectomy Catheter5 models · 6 reports · 1 recall
- Pruitt Occlusion Catheter6 models · 5 reports · 1 recall
- Pruitt Irrigation Occlusion Catheter2 models · 4 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page