XenoSure Biologic Patch 0.6BV8Recalls & reported problems
Lemaitre Vascular, Inc. · XenoSure Biologic Patch, 0.6cmx8cm · Class 2 · Implant
All models of XenoSure Biologic Patch (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840663106745), received between Apr 10, 2024 and Mar 6, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Compatibility Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Compatibility Problem4
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Inflammation4
- Obstruction/Occlusion4
- Stroke/CVA1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT HAD A CAROTID OPERATION THAT TRIGGERED A POST-OPERATIVE STROKE, WHICH REQUIRED RE-INTERVENTION. THE SURGEON SUSPECTS A POSSIBLE DEVICE FAILURE RELATED TO THE ADHESION OF THE DEVICE OF THE IMPLANTED DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED TO THE PATIENT. ADDITIONAL INFORMATION INCLUDING THE PRECISE IMPLANT DATE, DETAILS OF THE PROCEDURE AND INTERVENTION WERE REQUESTED. ADDITIONAL INFORMATION WAS ALSO REQUESTED RELATED TO THE ADDITIONAL ISSUES THAT HAVE OCCURRED, BUT NO ADDITIONAL INFORMATION HAS BEEN MADE AVAILABLE AT THIS TIME.
Patient problems: Stroke/CVA
MAUDE report 21533523 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085425 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085424 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085421 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085420 ↗
Other Mesh, Surgical devices
- Peri-Guard PC1016NSynovis Life Technologies, Inc. · Mesh, surgical1 recall1 report
- MEDPOR TITAN 81032Stryker Leibinger GmbH & Co. KG · Mesh, surgical1 recall1 report
- SurgiMend® 606-300-006Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-300-019Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-400-009Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-300-020Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-003-001Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-003-002Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page