XenoSure Biologic PatchRecalls & reported problems across all models
Lemaitre Vascular, Inc. · 7 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 22
- Recalls
- 0
We found 7 models of XenoSure Biologic Patch by Lemaitre Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 26 adverse event reports in the FDA's MAUDE database, received between Dec 8, 2017 and Apr 28, 2026 — classified as 22 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported XenoSure Biologic Patch problems
Device problems
- Patient Device Interaction Problem8
- Adverse Event Without Identified Device or Use Problem4
- Compatibility Problem4
- Insufficient Device Problem Information4
- Contamination1
- Contamination /Decontamination Problem1
- Delamination1
- Fluid/Blood Leak1
Patient problems
- Inflammation4
- Medical device site infection4
- Obstruction/Occlusion4
- Swelling4
- Bacterial Infection3
- Thrombosis3
- Abscess2
- Erythema2
XenoSure Biologic Patch models
7 models, most reported first. Each model has its own page with its reports, recalls and device details.
- XenoSure Biologic Patch e1P6Mesh, Surgical10 reports
- XenoSure Biologic Patch e0.8P8Mesh, Surgical7 reports
- XenoSure Biologic Patch 0.6BV8Mesh, Surgical5 reports
- XenoSure Biologic Patch e2P9Mesh, Surgical1 report
- XenoSure Biologic Patch e4P4Mesh, Surgical1 report
- XenoSure Biologic Patch e10P16Mesh, Surgical1 report
- XenoSure Biologic Patch e6P8Mesh, Surgical1 report
Recent XenoSure Biologic Patch adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryApr 28, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT THE XENOSURE TESTED POSITIVE ON CULTURE FOR CUTIBACTERIUM ACNES. THE CAUSE WAS ATTRIBUTED TO POSSIBLE OR CONTAMINATION. THE PRODUCT WAS IMPLANTED. IT IS NOT KNOWN WHAT ACTIONS, IF ANY, WERE TAKEN TO TREAT OR RESOLVE THE INFECTION, AS SPECIFIC DETAILS OF THE EVENT WERE NOT PROVIDED. NO OTHER PATIENT INJURY OR COMPLICATIONS WAS REPORTED.
Patient problems: Medical device site infection
MAUDE report 25016969 ↗InjuryFeb 5, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT A PATHOLOGY EXAMINATION OF THE PATIENT'S CAROTID PLAQUE TISSUE REVEALED ORGANISMS CONSISTENT WITH ASPERGILLUS. INFECTION CONTROL INDICATED THAT THIS FINDING MAY BE ASSOCIATED WITH THE BOVINE PATCH USED DURING A PRIOR CAROTID SURGERY. NO OTHER COMPLICATIONS OR PATIENT INJURY WAS REPORTED.
Patient problems: Medical device site infection
MAUDE report 24266852 ↗InjuryMay 29, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED A STAPH INFECTION FOLLOWING IMPLANTATION OF A 2X9 XENOSURE BIOLOGIC PATCH. ORIGINAL PROCEDURE WAS ON (B)(6) 2025 FOR A FEMORAL ENDARTERECTOMY WITH PATCH ANGIOPLASTY AND STENTING. THE PATIENT RETURNED ON (B)(6) 2025 FOR AN INCISION AND DRAINAGE OF THE WOUND. THE PATIENT WAS DISCHARGED HOME WITH A WOUND VAC AND THE WOUND IS IMPROVING. THE PATIENT IS STILL ON ANTIBIOTICS. THE PATCH REMAINS IMPLANTED.
Patient problems: Medical device site infection
MAUDE report 22107422 ↗MalfunctionMay 6, 2025Contamination; Contamination
IT WAS REPORTED THAT THE PATCH WAS IMPLANTED DURING A HECTIC CODE PERFUSION CRITICAL LAPAROTOMY (POST KIDNEY TRANSPLANT) WHERE PATIENT PRESENTED IN HEMORRHAGIC SHOCK PRE-OP. A SAMPLE OF THE GRAFT WAS SENT AFTER BEING OPENED (AND ADMITTEDLY HANDLED BY MULTIPLE PEOPLE; IN CONTACT WITH PATIENT AND PATIENT CONTAMINATED INSTRUMENTS/GLOVES). SAID SAMPLE CAME BACK POSITIVE FOR CULTURED OF CANDIDA FUNGUS. GRAFT WAS ALREADY IMPLANTED. NO UNUSUAL HARM OR INJURY HAS BEEN CAUSED BY THE INCIDENT AND PATIENT HAS SINCE BEEN DISCHARGED.
Patient problems: Bacterial Infection
MAUDE report 21968530 ↗InjuryMar 28, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT HAD A CAROTID ENDARTERECTOMY WITH PATCH ANGIOPLASTY USING THE XENOSURE PATCH. THE INITIAL SURGERY WAS PERFORMED ON (B)(6) 2025. THE PATIENT CAME BACK TO THE HOSPITAL WITH AN INFECTION ON (B)(6) 2025. ON (B)(6) 2025 THE INFECTION WAS REMOVED AND THE INTERVENTION WAS REPAIRED WITH THE PATIENT'S GREAT SAPHENOUS VEIN. THE VASCULAR COORDINATOR DID NOT KNOW THE EXACT CAUSE OF THE INFECTION. ADDITIONAL INFORMATION RECEIVED STATES THE CULTURE GREW STREP ANGINOUSUS.
Patient problems: Medical device site infection
MAUDE report 21718898 ↗InjuryMar 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT HAD A CAROTID OPERATION THAT TRIGGERED A POST-OPERATIVE STROKE, WHICH REQUIRED RE-INTERVENTION. THE SURGEON SUSPECTS A POSSIBLE DEVICE FAILURE RELATED TO THE ADHESION OF THE DEVICE OF THE IMPLANTED DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED TO THE PATIENT. ADDITIONAL INFORMATION INCLUDING THE PRECISE IMPLANT DATE, DETAILS OF THE PROCEDURE AND INTERVENTION WERE REQUESTED. ADDITIONAL INFORMATION WAS ALSO REQUESTED RELATED TO THE ADDITIONAL ISSUES THAT HAVE OCCURRED, BUT NO ADDITIONAL INFORMATION HAS BEEN MADE AVAILABLE AT THIS TIME.
Patient problems: Stroke/CVA
MAUDE report 21533523 ↗InjuryMay 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT POST CAROTID ENDARTERECTOMY WITH XENOSURE PATCH ANGIOPLASTY, THE PATIENT CAME BACK WITH A STAPHYLOCOCCUS INFECTION ORIGINATING FROM THE PATCH SITE. THE INFECTED PATCH WAS REMOVED AND THE PATIENT'S CAROTID WAS REPAIRED WITH APPLICATION OF NATIVE GSV GRAFT.
Patient problems: Bacterial Infection
MAUDE report 19322843 ↗MalfunctionMay 1, 2024
OPENED IMPLANT AND FOUND UNEXPECTED CLEAR PLASTIC FLOATING IN THE CONTAINER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220331 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085425 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085424 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085421 ↗InjuryApr 10, 2024Compatibility Problem; Compatibility Problem
THE DISTRIBUTOR WAS INFORMED BY THE CUSTOMER OF OBSTRUCTION CASES AFTER A FEW MONTHS OF XENOSURE IMPLANTATION: THERE HAVE BEEN FOUR CASES IN TOTAL, TWO OF THE SURGERIES WERE PERFORMED IN (B)(6) 2022 AND ANOTHER TWO IN (B)(6) 2023. THE SIZE USED WAS 0.6X0.8. ACCORDING TO OUR SUPPLY DATA, THE PRODUCT IMPLEMENTED IS IN ACCORDANCE WITH THE CODE 0.6BV8. AFTER A FEW MONTHS OF FOLLOW-UP AND UNDER ULTRASOUND, A SMALL OBSTRUCTION COULD BE OBSERVED IN ALL FOUR CASES. AFTER 8 MONTHS THIS OBSTRUCTION HAS INCREASED IN VOLUME, NOT YET BEING A CRITICAL OBSTRUCTION. THE OBSTRUCTION CANNOT BE DEFINED AS A COMMON PLAQUE, ITS TEXTURE IS SMOOTHER, AND ITS SHAPE IS LIKE AN HOURGLASS, AS IF IT WERE AN INFLAMMATORY REACTION. THE SURGEON HAS BEEN USING THIS PRODUCT SINCE 2019 AND HAS NOT HAD ANY CASES OF OBSTRUCTION SO FAR. ACCORDING TO THE INDICATIONS OF THE PATCH, IT IS NOT SPECIFIED ON WHICH SIDE IT IS RECOMMENDED TO IMPLANT THE PATCH, BUT THE SURGEON OBSERVED UNDER A MAGNIFYING GLASS THAT ONE OF THE SIDES IS FLATTER AND THIS IS THE SIDE THAT IMPLANT INWARDS. THE SURGEON IS NOT COMPLETELY CONVINCED THAT THE OBSTRUCTION IS CAUSED BY THE PATCH; HOWEVER, HE WANTED TO CONFIRM BY REQUESTING AN ANSWER FROM L
Patient problems: Inflammation; Obstruction/Occlusion
MAUDE report 19085420 ↗MalfunctionOct 13, 2023Material Discolored; Fluid/Blood Leak; Material Too Rigid or Stiff; Material Discolored; Fluid/Blood Leak; Material Too Rigid or Stiff
IT WAS REPORTED THAT THE XENOSURE BIOLOGIC PATCH THAT WAS USED DURING A CAROTID ENDARTERECTOMY PROCEDURE WAS DIFFERENT IN COLORING, TEXTURE AND FELT DIFFERENT (WOULDN'T CONTOUR, FELT HARD/LEATHERY). THE SURGEON CHOSE TO USE THE PATCH AND NOT REPLACE THE DEVICE WITH ANOTHER ONE. IT WAS REPORTED THAT AS A RESULT OF THE PATCH BEING STIFFER AND FEELING DIFFERENT, THE DEVICE TOOK LONGER TO SEW THAN NORMAL WHICH RESULTED IN LEAKAGE FROM THE SITE DURING THE PROCEDURE. ALTHOUGH THE PROCEDURE WAS DELAYED, THE SURGEON WAS SUCCESSFUL IN COMPLETING THE PROCEDURE USING THE DEVICE. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17930588 ↗InjurySep 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING LITERATURE REVIEW THE ARTICLE INDICATED THAT ONE LIMB (2%) WITHOUT POSTOPERATIVE WOUND COMPLICATIONS REQUIRED ADDITIONAL TREATMENT DUE TO ACUTE BLEEDING DURING A STUDY TITLED "CLINICAL OUTCOMES OF COMMON FEMORAL THROMBOENDARTERECTOMY WITH BOVINE PERICARDIUM PATCH ANGIOPLASTY".
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17698111 ↗InjuryJun 24, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
PATIENT HAD A CAROTID ENDARTERECTOMY WITH PATCH DONE ON 2020. SHE WAS PRESENTED IN 2022 WITH AN INFECTED PATCH REQUIRING SURGICAL INTERVENTION. FDA NOTIFIED THIS INCIDENT TO LEMAITRE ON 5/27/2022. FDA PROVIDED LEMAITRE WITH A COPY OF VOLUNTARY MEDWATCH REPORT# MW5109514.
Patient problems: Unspecified Infection
MAUDE report 14799415 ↗
Other Lemaitre Vascular, Inc. products
- LeMaitre Embolectomy Catheter47 models · 438 reports · 2 recalls
- Pruitt F3 Carotid Shunt10 models · 305 reports
- Artegraft Vascular Graft15 models · 25 reports · 3 recalls
- Trivex System3 models · 69 reports
- Syntel Silicone OTW Embolectomy Catheter8 models · 66 reports
- Syntel Silicone Thrombectomy Catheter5 models · 6 reports · 1 recall
- Pruitt Occlusion Catheter6 models · 5 reports · 1 recall
- Pruitt Irrigation Occlusion Catheter2 models · 4 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.