Skip to content

Syntel Silicone Thrombectomy Catheter A4545Recalls & reported problems

Lemaitre Vascular, Inc. · Syntel Silicone Thrombectomy Catheter · Class 2

All models of Syntel Silicone Thrombectomy Catheter (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840663109807), received between Mar 3, 2023 and Aug 30, 2023 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Apr 17, 2024Class IIOpen, Classified

    The guide tip can become damaged and result in the tip detaching.

    Z-1824-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Material Separation2
  • Device Slipped1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2023Material Separation; Material Separation

    IT WAS REPORTED SYNTEL 5F THROMBECTOMY CATHETER HAS A METAL SPRING TIP ON THE END WITH A SILICONE CAP, THIS CAP CAN BE TAKEN OFF WITH FINGERS DURING GRAFT/NATIVE FISTULA DECLOTTING PROCEDURE. WHEN INTERVENTIONIST ENCOUNTERS FRICTION, FROM THE SHEATH OR THE VESSEL WALL, THIS CAP BECOMES LOOSE, AND THE CONCERN IS THAT IT CAN END IN THE PATIENT'S LUNGS. THE SILICONE CAP WAS PLACED BACK ON THE METAL SPRING TIP. NO INJURY OCCURRED TO PATIENT. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT. REPORT 1220948-2023-00138 WAS SUBMITTED FOR THE SECOND DEVICE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17650615 ↗
  • MalfunctionAug 30, 2023Material Separation; Material Separation

    IT WAS REPORTED SYNTEL 5F THROMBECTOMY CATHETER HAS A METAL SPRING TIP ON THE END WITH A SILICONE CAP, THIS CAP CAN BE TAKEN OFF WITH FINGERS DURING GRAFT/NATIVE FISTULA DECLOTTING PROCEDURE. WHEN INTERVENTIONIST ENCOUNTERS FRICTION, FROM THE SHEATH OR THE VESSEL WALL, THIS CAP BECOMES LOOSE, AND THE CONCERN IS THAT IT CAN END IN THE PATIENT'S LUNGS. THE SILICONE CAP WAS PLACED BACK ON THE METAL SPRING TIP. NO INJURY OCCURRED TO PATIENT. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT. REPORT 1220948-2023-00137 WAS SUBMITTED FOR THE FIRST DEVICE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17650614 ↗
  • MalfunctionMay 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED AT THE TIME OF PERFORMING A MANEUVER IN THE PROCEDURE, THE GUIDE TIP TENDS TO DETACH IN THE AREA WHERE THE BALLOON IS LOCATED. THERE WERE NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16902474 ↗
  • MalfunctionMay 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THREE TIP OF THE CATHETERS WERE DAMAGED DURING EXTRACTION OF THE THROMBUS. PART OF THE TIP TENDS TO DETACH. THERE WERE NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16862199 ↗
  • MalfunctionMay 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THREE TIP OF THE CATHETERS WERE DAMAGED DURING EXTRACTION OF THE THROMBUS. PART OF THE TIP TENDS TO DETACH. THERE WERE NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16862198 ↗
  • MalfunctionMar 3, 2023Device Slipped; Device Slipped

    IT WAS REPORTED THAT THE TIP OF THE SYNTEL SILICONE THROMBECTOMY CATHETER BROKE OFF DURING THE THROMBECTOMY PROCEDURE. NO INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16477910 ↗

FDA clearance and classification

Cleared through 510(k) K910372 on Apr 26, 1991 (decision: Substantially Equivalent). Submitted by Applied Vascular Devices, Inc. as “Intimax Thrombectomy Catheter”. The FDA review took 87 days. FDA record ↗

The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Syntel Silicone Thrombectomy Catheter A4545

Has the Syntel Silicone Thrombectomy Catheter A4545 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1824-2024 (Class II), began on Apr 17, 2024: The guide tip can become damaged and result in the tip detaching. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Syntel Silicone Thrombectomy Catheter A4545?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Mar 3, 2023 and Aug 30, 2023. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (3), “Material Separation” (2) and “Device Slipped” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Syntel Silicone Thrombectomy Catheter A4545?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Syntel Silicone Thrombectomy Catheter A4545 FDA approved?

It was cleared through 510(k) K910372 on Apr 26, 1991. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Syntel Silicone Thrombectomy Catheter A4545, and is it still sold?

The FDA's device database lists Lemaitre Vascular, Inc. as the company responsible for it (brand Syntel Silicone Thrombectomy Catheter, model A4545). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page