Skip to content

Syntel Silicone OTW Embolectomy Catheter A4E09Recalls & reported problems

Lemaitre Vascular, Inc. · Syntel Silicone OTW Embolectomy Catheter · Class 2

All models of Syntel Silicone OTW Embolectomy Catheter (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840663109944), received between Aug 6, 2024 and May 13, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K970762); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture6

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Material Rupture

    IT WAS REPORTED THAT THE BALLOON WAS INFLATED TO PERFORM CLOT RETRIEVAL; HOWEVER, RESISTANCE WAS ENCOUNTERED DURING WITHDRAWAL FROM THE UPPER ARM PTFE GRAFT. DURING REMOVAL, THE BALLOON CATHETER BECAME PARTIALLY DISENGAGED FROM THE GUIDEWIRE AND SUBSEQUENTLY FRACTURED INTO TWO PIECES. THE DEVICE WAS ULTIMATELY REMOVED IN SEGMENTS USING A HEMOSTAT. NO ADDITIONAL DEVICE ISSUES OR PATIENT INJURY WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162855 ↗
  • MalfunctionAug 19, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE SYNTEL OVER-THE-WIRE EMBOLECTOMY CATHETER BALLOON RUPTURED DURING PRE-USE CHECK. IT WAS NOT USED ON THE PATIENT AND NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22846670 ↗
  • MalfunctionJul 8, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE SYNTEL OTW CATHETER BALLOON RUPTURED DURING A BRTO PROCEDURE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE OPERATION. THE DEVICE WAS CHECKED PRIOR TO USE. THE SURGEON SAW THE BALLOON RUPTURED THROUGH IMAGING. NO REPORT OF INJURY TO THE PATIENT. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22437724 ↗
  • MalfunctionJul 8, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE SYNTEL OTW CATHETER BALLOON RUPTURED DURING A BRTO PROCEDURE. THE SURGEON OVER INFLATED THE BALLOON. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE OPERATION. THE DEVICE WAS CHECKED PRIOR TO USE. THE SURGEON SAW THE BALLOON RUPTURED THROUGH IMAGING. NO REPORT OF INJURY TO THE PATIENT. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22437706 ↗
  • MalfunctionAug 6, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE SYNTEL OTW BALLOON RUPTURED. THEY WERE USING THE PYTHON BALLOON DURING AN OVARIAN VEIN EMBOLIZATION PROCEDURE AND UPON INFLATION THEY COULDN'T SEE ANY CONTRAST EVEN THOUGH THEY PUT SOME IN THE SYRINGE. DR. BABCOCK ASPIRATED THE BALLOON AND NOTICED IT HAD RUPTURED. THIS HAPPENED WITH TWO OF THE PYTHON BALLOONS. NO MORE THAN 2CC WAS INFLATED. THEY CHECKED THE INTEGRITY OF THE SECOND BALLOON (BOTH DEVICES PRE-CHECKED) BEFORE PLACING INSIDE THE PATIENT. IT WAS FINE UNTIL INFLATED INSIDE THE PATIENT, IT RUPTURED AGAIN. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT. REPORT 1220948-2024-00135 WAS SUBMITTED FOR THE FIRST DEVICE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19918710 ↗
  • MalfunctionAug 6, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE SYNTEL OTW BALLOON RUPTURED. THEY WERE USING THE PYTHON BALLOON DURING AN OVARIAN VEIN EMBOLIZATION PROCEDURE AND UPON INFLATION THEY COULDN'T SEE ANY CONTRAST EVEN THOUGH THEY PUT SOME IN THE SYRINGE. DR. BABCOCK ASPIRATED THE BALLOON AND NOTICED IT HAD RUPTURED. THIS HAPPENED WITH TWO OF THE PYTHON BALLOONS. NO MORE THAN 2CC WAS INFLATED. THEY CHECKED THE INTEGRITY OF THE SECOND BALLOON (BOTH DEVICES PRE-CHECKED) BEFORE PLACING INSIDE THE PATIENT. IT WAS FINE UNTIL INFLATED INSIDE THE PATIENT, IT RUPTURED AGAIN. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT. REPORT 1220948-2024-00147 WAS SUBMITTED FOR THE SECOND DEVICE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19918673 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K970762. One submission often covers many models, so check the product and lot details in each record.

  • Z-1846-2020 · Oct 24, 2019 · Terminated
    Python Embolectomy Catheters, Sterile EO, CE, Ref numbers/UDI: A4E01/(01)00607915113407; A4E02/(01)00607915113414; A4E03/(01)00607915113421; A4E04/(01)00607915113438; A4E05/(01
  • Z-1847-2020 · Oct 24, 2019 · Terminated
    Bard Embolectomy Catheter & 1 ml SYRINGE, REF numbers/UDI: CE0340DR/(01)H699CE0340DR10R; CE0380DR/(01)H699CE0380DR10V; CE0440DR/(01)H699CE0440DR10S; CE0480DR/(01)H699CE0480DR10

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page