Avance CS2Recalls & reported problems
Datex-Ohmeda Inc · AVANCE CS2 · Class 2
5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Avance CS2 is an anesthesia gas machine. Its 1,553 reports describe many cases of the machine failing to run on battery, low gas flow, and power or mechanical problems. The reports include 5 injuries and no deaths. There are 6 recalls, 5 still open, about a breathing tube that can be connected to the wrong port, backup batteries that can fail early, and a base that can crack.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,540
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 6
Summary
The FDA's MAUDE database contains 1,540 adverse event reports that list this exact device (DI 00840682102322), received between Feb 2, 2018 and Sep 30, 2026 — classified as 5 injuries, 1,535 malfunctions.
The most frequently reported specific device problem is “Failure to Run on Battery” (241 reports).
6 FDA recall records name this device's identifier, 5 still open.
23 other recalls cover devices cleared under the same FDA submission (K123125, K131945, K172045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 12, 2024Open, Classified
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Z-0078-2025 · Root cause: Device Design · FDA record ↗
- Jul 12, 2024Open, Classified
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Z-0079-2025 · Root cause: Device Design · FDA record ↗
- Jul 12, 2024Open, Classified
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Z-0080-2025 · Root cause: Device Design · FDA record ↗
- Jun 1, 2022Terminated
There is a potential reversal of the O2 and air cylinder pressure transducer connections.
Z-1389-2022 · Root cause: Process control · FDA record ↗
- May 27, 2022Open, Classified
The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.
Z-1427-2022 · Root cause: Process control · FDA record ↗
- Apr 25, 2022Open, Classified
Backup batteries can fail earlier than their estimated life.
Z-1285-2022 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Failure to Run on Battery241
- Insufficient Flow or Under Infusion229
- Therapeutic or Diagnostic Output Failure187
- Mechanical Problem158
- Power Problem141
- Electrical /Electronic Property Problem120
- Unexpected Therapeutic Results96
- Increase in Pressure68
Patient problems
- No Clinical Signs, Symptoms or Conditions1,478
- No Patient Involvement37
- No Known Impact Or Consequence To Patient11
- No Consequences Or Impact To Patient8
- Low Oxygen Saturation4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Cardiac Arrest1
- Coma1
Recent adverse event reports
The 25 most recent of 1,540 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Power Problem
It was reported that there was a malfunction during pre-use checkout resulting in power management system failure. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947553 ↗MalfunctionSep 30, 2026Failure to Run on Battery
It was reported that there was a battery failure during pre-use checkout that could cause loss of mechanical ventilation. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947134 ↗MalfunctionSep 29, 2026Failure to Run on Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941064 ↗MalfunctionSep 29, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934799 ↗MalfunctionSep 29, 2026Power Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933186 ↗MalfunctionSep 29, 2026Failure to Run on Battery
It was reported that there was a loss of mechanical ventilation during a case due to battery failure. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931661 ↗MalfunctionSep 29, 2026Mechanical Problem
It was reported that there was a malfunction resulting in inability to switch ventilation modes as expected during a case. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931588 ↗MalfunctionSep 29, 2026Insufficient Flow or Under Infusion
It was reported that there was a error resulting in insufficient O2 during pre-use checkout. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931447 ↗MalfunctionSep 28, 2026Unexpected Shutdown
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926615 ↗MalfunctionSep 28, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50925973 ↗MalfunctionSep 28, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50921974 ↗MalfunctionSep 25, 2026Failure to Run on Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905984 ↗MalfunctionSep 24, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889105 ↗MalfunctionSep 17, 2026Failure to Run on Battery
It was reported that there was a battery failure during service that could cause loss of mechanical ventilation. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806993 ↗MalfunctionAug 27, 2026Mechanical Problem
It was reported that there was a malfunction during pre-use checkout resulting in loss of mechanical ventilation. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50473758 ↗MalfunctionAug 17, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50276220 ↗MalfunctionAug 5, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147537 ↗MalfunctionAug 5, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50143012 ↗MalfunctionJul 1, 2026Failure to Charge
IT WAS REPORTED THAT THERE WAS A FAILURE TO CHARGE THE BATTERY THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25678074 ↗MalfunctionJul 1, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INABILITY TO SWITCH VENTILATION MODES AS EXPECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25675577 ↗MalfunctionJun 30, 2026Power Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN LOSS OF UNIT POWER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668313 ↗MalfunctionJun 30, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668299 ↗MalfunctionJun 30, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25664568 ↗MalfunctionJun 29, 2026Power Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF POWER AND SUBSEQUENT LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25658074 ↗MalfunctionJun 29, 2026Increase in Pressure
IT WAS REPORTED THAT THERE WAS AN MALFUNCTION RESULTING IN OVER DELIVERY OF PRESSURE DURING A CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25650159 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123125, K131945, K172045. One submission often covers many models, so check the product and lot details in each record.
- Z-3191-2026 · Jul 30, 2026 · Open, ClassifiedAisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
- Z-3190-2026 · Jul 30, 2026 · Open, ClassifiedAisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
- Z-3192-2026 · Jul 30, 2026 · Open, ClassifiedEt Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000
- Z-3189-2026 · Jul 30, 2026 · Open, ClassifiedAisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
- Z-0075-2025 · Jul 12, 2024 · Open, ClassifiedAespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals t
- Z-0082-2025 · Jul 12, 2024 · Open, ClassifiedAisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals tra
- Z-0081-2025 · Jul 12, 2024 · Open, ClassifiedAisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained
- Z-0076-2025 · Jul 12, 2024 · Open, ClassifiedAespire 7900. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals
- Z-0077-2025 · Jul 12, 2024 · Open, ClassifiedAespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals
- Z-0074-2025 · Jul 12, 2024 · Open, ClassifiedAespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory suppo
- Z-0084-2025 · Jul 12, 2024 · Open, ClassifiedAuxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, howev
- Z-1286-2022 · Apr 25, 2022 · Open, ClassifiedField replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices
- Z-2596-2021 · Aug 2, 2021 · Open, ClassifiedFlow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Care
- Z-2595-2021 · Aug 2, 2021 · Open, ClassifiedFlow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001 FLOW SENSOR, LEGACY VAR ORF BCG (blue, cleanable) 2089610-001-S FLOW
- Z-0314-2021 · Sep 17, 2020 · Open, Classified9100c NXT system - Product Usage: Device is licensed and distributed outside of the US only.
…and 8 more.
Other Gas-machine, anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,109 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,204 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
- Atlan A350 8621500Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia4 recalls63 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.