Carestation 750 A1Recalls & reported problems
Datex-Ohmeda Inc · CARESTATION 750 A1 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Carestation 750 anesthesia machine has 2,202 reports, with no deaths and 3 injuries. Most describe the machine delivering no gas flow, and others describe output failures, electrical problems or mechanical problems. 2 recalls are open: one over a possible unexpected shutdown during AC power loss, and one over poor ventilation in volume control mode on certain models. Reports fell to 705 in the last 12 months from 1,308 in the previous 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,193
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2,193 adverse event reports that list this exact device (DI 00840682145596), received between Oct 5, 2023 and Sep 29, 2026 — classified as 3 injuries, 2,190 malfunctions.
The most frequently reported specific device problem is “No Flow” (1,431 reports).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K213867); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 14, 2025Class IOpen, Classified
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Z-0914-2026 · Root cause: Component change control · Recall details · FDA record ↗
- Mar 21, 2025Class IOpen, Classified
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Z-1635-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Flow1,431
- Therapeutic or Diagnostic Output Failure124
- Insufficient Flow or Under Infusion115
- Electrical /Electronic Property Problem108
- Mechanical Problem96
- Unexpected Therapeutic Results82
- Failure to Run on Battery76
- Power Problem48
Patient problems
- No Clinical Signs, Symptoms or Conditions2,191
- Low Oxygen Saturation2
- Laceration(s)1
Recent adverse event reports
The 25 most recent of 2,193 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Insufficient Flow or Under Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932139 ↗MalfunctionSep 18, 2026Failure to Run on Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829657 ↗MalfunctionSep 10, 2026Insufficient Flow or Under Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714771 ↗MalfunctionSep 10, 2026Unexpected Therapeutic Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714768 ↗MalfunctionSep 8, 2026Insufficient Flow or Under Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50676799 ↗MalfunctionAug 27, 2026Failure to Run on Battery
It was reported that there was a battery failure that could cause loss of mechanical ventilation. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50484263 ↗MalfunctionAug 27, 2026Insufficient Flow or Under Infusion
It was reported there was a malfunction during a case resulting in insufficient oxygen or fresh gas flow. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50474707 ↗MalfunctionAug 20, 2026Improper Flow or Infusion; No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50355040 ↗MalfunctionAug 19, 2026Installation-Related Problem; No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50312362 ↗MalfunctionAug 19, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50312314 ↗MalfunctionAug 18, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50300516 ↗MalfunctionAug 17, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50288730 ↗MalfunctionAug 17, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50287340 ↗MalfunctionAug 17, 2026Unexpected Therapeutic Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50286732 ↗MalfunctionAug 17, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50280615 ↗MalfunctionJul 1, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25680412 ↗MalfunctionJul 1, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25680134 ↗MalfunctionJul 1, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25679825 ↗MalfunctionJun 30, 2026No Flow
THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE THAT INDICATES IT MAY BE IMPACTED BY A CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON (B)(6) 2025 (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025). THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25667489 ↗MalfunctionJun 29, 2026Failure to Power Up
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN POWER SUPPLY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25661922 ↗MalfunctionJun 29, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25658533 ↗MalfunctionJun 29, 2026Unexpected Therapeutic Results
IT WAS REPORTED THAT THERE WAS AN ERROR DURING PRE-USE CHECKOUT RESULTING IN UNEXPECTED THERAPEUTIC SYSTEM PERFORMANCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25651234 ↗MalfunctionJun 26, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN VENTILATION FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25642172 ↗MalfunctionJun 26, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN UNEXPECTED THERAPEUTIC SYSTEM PERFORMANCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636408 ↗MalfunctionJun 26, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN BATTERY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636155 ↗
FDA clearance and classification
Cleared through 510(k) K213867 on Mar 28, 2023 (decision: Substantially Equivalent). Submitted by Datex-Ohmeda, Inc. as “Carestation 750/750c”. The FDA review took 473 days. FDA record ↗
The FDA regulates product code BSZ (gas-machine, anesthesia) under 21 CFR 868.5160 as a Class II device. 21 CFR 868.5160 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Recalls under the same FDA submission
These recalls involve devices cleared under K213867. One submission often covers many models, so check the product and lot details in each record.
- Z-0915-2026 · Nov 14, 2025 · Open, ClassifiedGE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S
- Z-1637-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.
Common questions about the Carestation 750 A1
Has the Carestation 750 A1 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0914-2026 (Class I), began on Nov 14, 2025: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Carestation 750 A1?
The FDA's MAUDE database lists 2,193 adverse event reports for this device, received between Oct 5, 2023 and Sep 29, 2026. In those reports, the most common specific device problems are “No Flow” (1,431), “Therapeutic or Diagnostic Output Failure” (124) and “Insufficient Flow or Under Infusion” (115); the most common patient problems are “Low Oxygen Saturation” (2) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Carestation 750 A1?
None of the reports are classified as deaths, 3 as injuries and 2,190 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Carestation 750 A1 FDA approved?
It was cleared through 510(k) K213867 on Mar 28, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Carestation 750 A1 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Carestation 750 A1, and is it still sold?
The FDA's device database lists Datex-Ohmeda Inc as the company responsible for it (brand Carestation 750, model A1). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gas-machine, anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,073 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,029 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls1,977 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,535 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls17 reports
- Atlan A350 8621500Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia4 recalls61 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page