Carestation 650 A1Recalls & reported problems
Datex-Ohmeda Inc · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Carestation 650 is an anesthesia gas machine. Its 3,102 reports mostly describe no gas flow, failure to run on battery, and output and power failures. The reports include 1 injury. There are 2 open recalls, covering possible unexpected shutdowns and poor ventilation in some modes. Reports fell in the most recent year (862) from the year before (1,393).
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 3,109
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3,109 adverse event reports that list this exact device (DI 00840682103947), received between Feb 2, 2018 and Sep 30, 2026 — classified as 1 injury, 3,108 malfunctions.
The most frequently reported specific device problem is “No Flow” (980 reports).
2 FDA recall records name this device's identifier, 2 still open.
12 other recalls cover devices cleared under the same FDA submission (K151570); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 14, 2025Open, Classified
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Z-0913-2026 · Root cause: Component change control · FDA record ↗
- Mar 21, 2025Open, Classified
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Z-1628-2025 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- No Flow980
- Failure to Run on Battery427
- Therapeutic or Diagnostic Output Failure422
- Mechanical Problem254
- Power Problem221
- Electrical /Electronic Property Problem173
- Insufficient Flow or Under Infusion164
- Unexpected Therapeutic Results101
Patient problems
- No Clinical Signs, Symptoms or Conditions3,011
- No Patient Involvement69
- No Consequences Or Impact To Patient15
- No Known Impact Or Consequence To Patient6
- Patient Problem/Medical Problem2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Capsular Contracture1
- Chills1
Recent adverse event reports
The 25 most recent of 3,109 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Insufficient Flow or Under Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955025 ↗MalfunctionSep 30, 2026Failure to Run on Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954933 ↗MalfunctionSep 30, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949545 ↗MalfunctionSep 30, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948961 ↗MalfunctionSep 30, 2026Therapeutic or Diagnostic Output Failure
It was reported that there was an error during pre-use checkout resulting in loss of mechanical ventilation.There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945862 ↗MalfunctionSep 30, 2026No Flow
As a result of an inspection that was completed as part of the correction initiated by GE HealthCare (GEHC) on 02-APR-2025, (Recall numbers Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) this unit was identified as having an issue that prevents effective mechanical ventilation in Volume Control Ventilation (VCV) mode. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942991 ↗MalfunctionSep 30, 2026No Flow
As a result of an inspection that was completed as part of the correction initiated by GE HealthCare (GEHC) on 02-APR-2025, (Recall numbers Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) this unit was identified as having an issue that prevents effective mechanical ventilation in Volume Control Ventilation (VCV) mode. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942935 ↗MalfunctionSep 29, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940613 ↗MalfunctionSep 29, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932851 ↗MalfunctionSep 29, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931809 ↗MalfunctionSep 28, 2026Power Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926701 ↗MalfunctionSep 28, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917593 ↗MalfunctionSep 28, 2026Inappropriate or Unexpected Reset
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917537 ↗MalfunctionSep 28, 2026No Flow
As a result of an inspection that was completed as part of the correction initiated by GE HealthCare (GEHC) on 02-APR-2025, (Recall numbers Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) this unit was identified as having an issue that prevents effective mechanical ventilation in Volume Control Ventilation (VCV) mode. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50914004 ↗MalfunctionSep 28, 2026Excess Flow or Over-Infusion
It was reported that there was a malfunction resulting in high CO2 delivery during pre-use checkout. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913455 ↗MalfunctionSep 27, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910421 ↗MalfunctionSep 27, 2026Failure to Run on Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910410 ↗MalfunctionSep 27, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910409 ↗MalfunctionSep 25, 2026Power Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906078 ↗MalfunctionSep 25, 2026Unexpected Shutdown
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906032 ↗MalfunctionSep 25, 2026Therapeutic or Diagnostic Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50904389 ↗MalfunctionSep 25, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50902920 ↗MalfunctionSep 25, 2026Therapeutic or Diagnostic Output Failure
It was reported that there was a malfunction resulting in loss of mechanical ventilation during pre-use checkout. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895783 ↗MalfunctionSep 25, 2026Power Problem
It was reported that there was a malfunction resulting in power up problem causing loss of ventilation during pre-use checkout. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895771 ↗MalfunctionSep 25, 2026Therapeutic or Diagnostic Output Failure
It was reported that there was a malfunction resulting in loss of mechanical ventilation. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895587 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151570. One submission often covers many models, so check the product and lot details in each record.
- Z-0915-2026 · Nov 14, 2025 · Open, ClassifiedGE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S
- Z-1632-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.
- Z-1624-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems
- Z-1627-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
- Z-1634-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.
- Z-1631-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
- Z-0813-2020 · Nov 25, 2019 · TerminatedCarestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric
- Z-2954-2018 · Jul 24, 2018 · TerminatedGE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems
- Z-2955-2018 · Jul 24, 2018 · TerminatedGE Carestation 30 anesthesia systems
- Z-2956-2018 · Jul 24, 2018 · TerminatedGE 9100c NXT anesthesia systems
- Z-0579-2018 · Oct 11, 2017 · Open, ClassifiedCarestation 620 A1, Carestation 650 A1, and Carestation 650c A1
- Z-2703-2017 · May 2, 2017 · TerminatedCarestation 620 A1, Carestation 650 A1, and Carestation 650c A1
Other Gas-machine, anesthesia devices
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,204 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,540 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
- Atlan A350 8621500Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia4 recalls63 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.