NA SIGNA Creator (UDI 00840682113786)Recalls & reported problems
Ge Healthcare(Tianjin) Company Limited · SIGNA Creator System Medical Device · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682113786), received between Feb 25, 2025 and Feb 25, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
13 other recalls cover devices cleared under the same FDA submission (K143251); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 31, 2025Open, Classified
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Z-1239-2025 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Use of Device Problem1
Patient problems
- Laceration(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 25, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
DURING SERVICING OF A GEHC MR SYSTEM, A SERVICE CONTRACTOR BROUGHT A FERROUS CRIMPING TOOL INTO THE MAGNET ROOM WHEN IT WAS PULLED FROM THEIR HAND BY THE MAGNET. IT IS REPORTED THE SERVICE CONTRACTOR SUSTAINED A DISLOCATED FINGER AND LACERATION TO THE RIGHT PALM THAT WAS SUTURED.
Patient problems: Laceration(s)
MAUDE report 21458705 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143251. One submission often covers many models, so check the product and lot details in each record.
- Z-1242-2025 · Jan 31, 2025 · Open, ClassifiedSIGNA MR355, SIGNA MR360, NMRI system
- Z-2475-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA MR355, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
- Z-2473-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2476-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
- Z-2472-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA Creator, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2470-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA Voyager, with affected software versions: PX26.0 to PX26.6, VX28.0, VX29.1, VX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-0597-2022 · Dec 23, 2021 · Open, ClassifiedGE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA MR355 and SIGNA MR360 System, nuclear magnetic resonance imaging system.
- Z-0596-2022 · Dec 23, 2021 · Open, ClassifiedGE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Creator and SIGNA Explorer System, nuclear magnetic resonance imaging system.
- Z-0137-2022 · Aug 26, 2021 · TerminatedSIGNA Creator, SIGNA Explorer nuclear magnetic resonance imaging system
- Z-2058-2021 · Jun 4, 2021 · Open, ClassifiedSIGNA Creator, SIGNA Explorer, Nuclear Magnetic Resonance Imaging System
- Z-2059-2021 · Jun 4, 2021 · Open, ClassifiedSIGNA MR355, SIGNA MR360, Nuclear Magnetic Resonance Imaging System
- Z-3026-2017 · May 11, 2017 · TerminatedSIGNA Creator, SIGNA ExplorerProduct Usage: 1.5T SIGNA Creator and 1.5T SIGNA Explorer is a whole body magnetic resonance scanner designed to support high resolution, high signal-t
- Z-2731-2016 · Aug 31, 2016 · TerminatedGE Healthcare SIGNA Creator, SIGNA Explorer. Diagnostic Imaging Device.
Other System, nuclear magnetic resonance imaging devices
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
- Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging13 recalls2 reports
- NA Discovery MR750w 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging13 recalls0 reports
- Ingenia Ingenia Ambition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging12 recalls12 reports
- Ingenia Ingenia 3.0T CXPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls59 reports
- Ingenia Smartpath to dStream for 3.0TPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls44 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls5 reports
- MR 7000 Series MR 7700Philips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page