Skip to content

NA SIGNA Artist (UDI 00840682123129)Recalls & reported problems

GE Medical Systems, L.L.C. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
5

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682123129), received between Nov 8, 2024 and Nov 8, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

5 FDA recall records name this device's identifier, 5 still open.

36 other recalls cover devices cleared under the same FDA submission (K160618); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 20, 2025Class IIOpen, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2299-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

  2. Jun 20, 2025Class IIOpen, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2303-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

  3. Jun 20, 2025Class IIOpen, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2306-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

  4. Jun 20, 2025Class IIOpen, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2310-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

  5. Jun 20, 2025Class IIOpen, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2302-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT FELT "INTENSE AND STRONG HEAT" ON THEIR RIGHT SIDE DURING A MR EXAMINATION OF THE RIGHT SHOULDER. THERE IS NO REPORT OF A BURN OR OTHER TYPICAL SIGN OF SIGNIFICANT RF-RELATED INJURY. THIS EVENT WAS INITIALLY ASSESSED AS NOT REPORTABLE AS WARMING DURING AN MR EXAM MAY OCCUR DUE TO THE RF ENERGY, AND TYPICALLY RESULTS IN MINOR DISCOMFORT WHICH IS SELF-RESOLVING. SUBSEQUENTLY, THE GE HEALTHCARE COMPLAINT HANDLING UNIT BECAME AWARE OF ADDITIONAL INFORMATION FROM THE PATIENT THAT DESCRIBES ONGOING PAIN THAT REQUIRES PAIN MEDICATION; THEREFORE, WE HAVE UPDATED OUR ASSESSMENT TO A SERIOUS INJURY. THE CAUSE OR SOURCE OF THE ONGOING PAIN HAS NOT BEEN CLEARLY IDENTIFIED. BASED ON THE SERVICE AND PREVENTATIVE MAINTENANCE RECORDS THAT WE HAVE REVIEWED FOR THE SYSTEM FROM THE TIME OF THE EVENT UNTIL NOW, WE DO NOT HAVE ANY EVIDENCE THAT THE SYSTEM MALFUNCTIONED, HOWEVER OUR INVESTIGATION IS ONGOING.

    Patient problems: Pain

    MAUDE report 20643359 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160618. One submission often covers many models, so check the product and lot details in each record.

  • Z-2301-2025 · Jun 20, 2025 · Open, Classified
    GE Discovery MR450 1.5T, Nuclear Magnetic Resonance Imaging System
  • Z-2130-2025 · May 23, 2025 · Open, Classified
    SIGNA Architect
  • Z-0518-2025 · Oct 28, 2024 · Open, Classified
    SIGNA Architect, Nuclear Magnetic Resonance Imaging System
  • Z-0519-2025 · Oct 28, 2024 · Open, Classified
    SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System
  • Z-0517-2025 · Oct 28, 2024 · Open, Classified
    Discovery MR750w 3.0T, whole body magnetic resonance scanner
  • Z-2459-2024 · Jun 14, 2024 · Open, Classified
    SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging
  • Z-2467-2024 · Jun 14, 2024 · Open, Classified
    Discovery MR450 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0; Nuclear Magnetic Resonance Imaging System
  • Z-2465-2024 · Jun 14, 2024 · Open, Classified
    Discovery MR750w 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.5, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
  • Z-2464-2024 · Jun 14, 2024 · Open, Classified
    Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
  • Z-2468-2024 · Jun 14, 2024 · Open, Classified
    Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
  • Z-2469-2024 · Jun 14, 2024 · Open, Classified
    SIGNA Artist, with affected software versions: DV26.0, DV27.1, DV28.1, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
  • Z-0593-2022 · Dec 23, 2021 · Open, Classified
    GE Healthcare MR superconducting magnets, a component of GE Healthcare Discovery MR750w 3.0T System, nuclear magnetic resonance imaging system.
  • Z-0591-2022 · Dec 23, 2021 · Open, Classified
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Architect system, nuclear magnetic resonance imaging system.
  • Z-0602-2022 · Dec 23, 2021 · Open, Classified
    GE Healthcare MR superconducting magnets, a component of GE Healthcare Discovery MR450 1.5T system, nuclear magnetic resonance imaging system.
  • Z-0594-2022 · Dec 23, 2021 · Open, Classified
    GE Healthcare MR superconducting magnets, a component of GE Healthcare Discovery MR750 3.0T System, nuclear magnetic resonance imaging system.

…and 21 more.

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page