Skip to content

CombiSet 03-2621-5Recalls & reported problems

Fresenius Medical Care Holdings, Inc. · Post-pump, 8.0 mm · Class 2

All models of CombiSet (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840861100224), received between Jan 15, 2020 and Jan 29, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

2 other recalls cover devices cleared under the same FDA submission (K962081); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak4

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Anemia1
  • Chest Pain1
  • Dyspnea1
  • Hemorrhage/Blood Loss/Bleeding1
  • Weakness1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    A USER FACILITY¿S TECHNICAL MANAGER REPORTED THAT A COMBISET BLOODLINE LEAKED ABOUT THIRTY MINUTES INTO A PATIENT¿S HEMODIALYSIS (HD) TREATMENT. THE PATIENT¿S ESTIMATED BLOOD LOSS (EBL) IS UNKNOWN. IT IS UNKNOWN IF THE COMPLAINT DEVICE IS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. MULTIPLE ATTEMPTS HAVE BEEN MADE TO CONTACT THE CUSTOMER TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE FILE WILL BE UPDATED ACCORDINGLY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21263702 ↗
  • MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A USER FACILITY BIOMEDICAL TECHNICIAN (BIOMED) REPORTED TO FRESENIUS CUSTOMER SERVICE THAT A COMBI SET BLOOD LEAK OCCURRED DURING A PATIENT¿S HEMODIALYSIS (HD) TREATMENT. ADDITIONAL INFORMATION WAS OBTAINED THROUGH FOLLOW-UP. ABOUT AN HOUR INTO THE PATIENT¿S HD TREATMENT, THE HEPARIN LINE ¿BLEW OFF¿ (OR SEPARATED) AT THE RED ¿T¿ CONNECTION. THERE WERE NO MACHINE ALARMS. THE PATIENT WAS DIALYZING ON A FRESENIUS 2008T HD MACHINE AND UTILIZING AN ELISIO DIALYZER. THERE WERE NO VISIBLE DEFECTS NOTED ON THE BLOODLINE DURING THE PRE-TREATMENT INSPECTION. IT WAS ALSO CONFIRMED THAT THERE WAS NO LEAKING DURING THE PRIMING. THE PATIENT¿S BLOOD WAS NOT RETURNED; ESTIMATED BLOOD LOSS (EBL) WAS REPORTEDLY 250 ML. IT WAS CONFIRMED THERE WAS NO PATIENT INJURY, NO ADVERSE EFFECTS WERE EXPERIENCED, AND NO MEDICAL INTERVENTION WAS REQUIRED AS A RESULT OF THE REPORTED EVENT. THE PATIENT COMPLETED THEIR TREATMENT AFTER BEING RE-SETUP WITH NEW SUPPLIES ON THE SAME MACHINE. THE SAMPLE WAS REPORTED TO BE AVAILABLE FOR MANUFACTURER EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15571201 ↗
  • MalfunctionOct 10, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A USER FACILITY BIOMEDICAL TECHNICIAN (BIOMED) REPORTED TO FRESENIUS CUSTOMER SERVICE THAT A COMBI SET BLOOD LEAK OCCURRED DURING A PATIENT¿S HEMODIALYSIS (HD) TREATMENT. ADDITIONAL INFORMATION WAS OBTAINED THROUGH FOLLOW-UP. IMMEDIATELY AFTER THE PATIENT WAS PUT ON THE MACHINE FOR HD TREATMENT, THE NURSE OBSERVED BLOOD DRIPPING FROM THE HEPARIN LINE RED ¿T¿ CONNECTION. THERE WERE NO MACHINE ALARMS. THE PATIENT WAS DIALYZING ON A FRESENIUS 2008T HD MACHINE AND UTILIZING AN ELISIO DIALYZER. THERE WERE NO VISIBLE DEFECTS NOTED ON THE BLOODLINE DURING THE PRE-TREATMENT INSPECTION. IT WAS ALSO CONFIRMED THAT THERE WAS NO LEAKING DURING THE PRIMING. THE PATIENT¿S BLOOD LOSS WAS MINIMAL; THEIR ESTIMATED BLOOD LOSS (EBL) WAS 2 ML. IT WAS CONFIRMED THERE WAS NO PATIENT INJURY, NO ADVERSE EFFECTS WERE EXPERIENCED, AND NO MEDICAL INTERVENTION WAS REQUIRED AS A RESULT OF THE REPORTED EVENT. THE PATIENT COMPLETED THEIR TREATMENT AFTER BEING RE-SETUP WITH NEW SUPPLIES ON THE SAME MACHINE. THE SAMPLE WAS REPORTED TO BE AVAILABLE FOR MANUFACTURER EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15569870 ↗
  • InjuryJan 15, 2020Fluid/Blood Leak; Fluid/Blood Leak

    A HEMODIALYSIS (HD) OUTPATIENT CLINIC REPORTED THAT A BLOOD LEAK OCCURRED AN HOUR AND FIFTEEN MINUTES INTO A PATIENT¿S HD TREATMENT ON (B)(6) 2019. THE STAFF NOTICED BLOOD LEAKING FROM BEHIND THE PATIENT¿S CHAIR. THE LEAK WAS REPORTEDLY COMING FROM A HOLE IN THE VENOUS LINE OF THE COMBI SET. THE PATIENT¿S ESTIMATED BLOOD LOSS (EBL) WAS 500 ML. AFTER THE BLOOD LEAK WAS DISCOVERED, THE TREATMENT WAS ENDED, THE BLOOD WAS CLEANED UP, AND THE PATIENT WAS TRANSFERRED TO A NEW MACHINE WITH STABLE VITAL SIGNS. AFTER COMPLETING TREATMENT ON THE NEW MACHINE, THE PATIENT WAS DISCHARGED IN STABLE CONDITION. THE PATIENT AND THEIR SPOUSE WERE EDUCATED ON SIGNS OF ANEMIA AND INSTRUCTED TO GO TO THE EMERGENCY ROOM (ER) SHOULD ANY SIGNS/SYMPTOMS APPEAR. ON (B)(6) 2019 THE PATIENT EXPERIENCED CHEST PAIN AND SHORTNESS OF BREATH. THE PATIENT WENT TO THE ER AND WAS PUT UNDER OBSERVATION. THE PATIENT¿S HEMOGLOBIN (HGB) ON ARRIVAL TO THE ER WAS 7.8 WHICH WAS A DROP FROM THE (B)(6) WHEN IT WAS 9.6. THE PATIENT RECEIVED TWO UNITS OF PACKED RED BLOOD CELLS (PRBC) AND WAS DISCHARGED TO THEIR HOME ON (B)(6) 2019 WITH A HGB OF 8.7. THE PATIENT RETURNED TO THE HD CLINIC ON (B)(6) 2019 AND HAS BEEN CONTINUING TR

    Patient problems: Anemia; Chest Pain; Dyspnea; Weakness

    MAUDE report 9594306 ↗

FDA clearance and classification

Cleared through 510(k) K962081 on Nov 1, 1996 (decision: Substantially Equivalent). Submitted by National Medical Care, Medical Products Div., Inc. as “Combilines Hemodialysis Blood Tubing Set”. The FDA review took 157 days. FDA record ↗

The FDA regulates product code FJK (set, tubing, blood, with and without anti-regurgitation valve) under 21 CFR 876.5820 as a Class II device. 21 CFR 876.5820 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Recalls under the same FDA submission

These recalls involve devices cleared under K962081. One submission often covers many models, so check the product and lot details in each record.

  • Z-0843-2011 · Nov 19, 2010 · Terminated
    Fresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2695-9
  • Z-0844-2011 · Nov 19, 2010 · Terminated
    Fresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2795-7

Common questions about the CombiSet 03-2621-5

Has the CombiSet 03-2621-5 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00840861100224), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the CombiSet 03-2621-5?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 15, 2020 and Jan 29, 2025. In those reports, the most common specific device problem is “Fluid/Blood Leak” (4); the most common patient problems are “Anemia” (1), “Chest Pain” (1) and “Dyspnea” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CombiSet 03-2621-5?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CombiSet 03-2621-5 FDA approved?

It was cleared through 510(k) K962081 on Nov 1, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CombiSet 03-2621-5, and is it still sold?

The FDA's device database lists Fresenius Medical Care Holdings, Inc. as the company responsible for it (brand CombiSet, model 03-2621-5). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices

All SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page