OVERWATCH 108-107-01-SRecalls & reported problems

Spinal Elements · Quick Lock Screw Driver, Short Inner Shaft · Class 1

All models of OVERWATCH (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840916134686), received between Jun 10, 2021 and May 8, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Mechanical Problem5

Patient problems

  • Foreign Body In Patient4
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Fracture; Mechanical Problem

    ON (B)(6) 2026 PATIENT UNDERWENT REVISION SURGERY TO REMOVE HARDWARE FOR SPINAL FUSION SURGERY FOR ADJACENT SEGMENT DISEASE AFTER SUCCESSFUL FUSION HAD OCCURRED AT INDEX LEVEL. REPORTEDLY DUE TO THE PATIENT'S SCLEROTIC BONE AND BONE INGROWTH, THE BONE SCREW WAS FUSED TO THE BONE. INTRA-OPERATIVELY EXCESSIVE FORCE WAS USE AND THE BONE SCREWDRIVER DEVICE FRACTURED AT THE TIP. THE SURGEON ELECTED TO LEAVE THE BONE SCREW AND THE DEVICE TIP ENCASED IN THE SCREW HEAD AND COMPLETED THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25124977 ↗
  • MalfunctionOct 12, 2021Fracture; Mechanical Problem; Fracture; Mechanical Problem

    WHILE DRIVING THE PEDICLE BONE SCREW, REPORTEDLY THE DRIVER TIP FRACTURED AND REMAINED IN THE HEAD OF THE SCREW SHANK. THOUGH THE DRIVER TIP WAS IDENTIFIED, THE SURGEON ELECTED TO LEAVE THE DRIVER TIP AND THE SCREW IN PLACE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12619060 ↗
  • MalfunctionOct 12, 2021Fracture; Mechanical Problem; Fracture; Mechanical Problem

    WHILE DRIVING THE PEDICLE BONE SCREW, REPORTEDLY THE DRIVER TIP FRACTURED AND REMAINED IN THE HEAD OF THE SCREW SHANK. THOUGH THE DRIVER TIP WAS IDENTIFIED, THE SURGEON ELECTED TO LEAVE THE DRIVER TIP AND THE SCREW IN PLACE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12619010 ↗
  • MalfunctionAug 2, 2021Fracture; Mechanical Problem; Fracture; Mechanical Problem

    WHILE DRIVING THE PEDICLE BONE SCREW, REPORTEDLY THE DRIVER TIP FRACTURED AND REMAINED IN THE HEAD OF THE RIGHT L5 SCREW SHANK. LEVELS BEING TREATED WERE T9--S2AI. THOUGH THE DRIVER TIP WAS IDENTIFIED, THE SURGEON ELECTED TO LEAVE THE DRIVER TIP AND THE SCREW IN PLACE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12264188 ↗
  • MalfunctionJun 10, 2021Fracture; Mechanical Problem; Fracture; Mechanical Problem

    WHILE DRIVING THE PEDICLE BONE SCREW, REPORTEDLY THE DRIVER TIP FRACTURED AND REMAINED IN THE HEAD OF THE SCREW SHANK. LEVELS BEING TREATED WERE L1-S1. THOUGH THE DRIVER TIP WAS IDENTIFIED, THE SURGEON ELECTED TO LEAVE THE DRIVER TIP AND THE SCREW IN PLACE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 11978438 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →