Skip to content

Arsenal 87107-055Recalls & reported problems

Alphatec Spine, Inc. · 5.5mm TAP · Class 1

All models of Arsenal (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840967126838), received between Feb 21, 2019 and Feb 11, 2020 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break5
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2020Break

    TAP WAS IN THE PEDICLE AND SURGEON FORCED DOWN AT AN ANGLE CAUSING IT TO BREAK. THE EVENT CAUSED NO PATIENT HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9694012 ↗
  • MalfunctionDec 10, 2019Break

    SURGEON WAS TAPPING INTO BONE AND IT SNAPPED RIGHT ABOVE THE THREADS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9444000 ↗
  • MalfunctionMay 29, 2019Break

    WHILE TAPPING THE PEDICLE, THE 5.5MM TAP BROKE. THE SURGEON REMOVED THE THREADED TIP FROM THE PATIENTS BONE AND THE SURGERY COMPLETED WITHOUT FURTHER ISSUE. THE EVENT CAUSED NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8649801 ↗
  • MalfunctionMar 1, 2019Break; Detachment of Device or Device Component

    WHILE TAPPING THE PEDICLE, THE 5.5MM TAP BROKE. THE DETACHED SECTION WAS RETRIEVED FROM THE PATIENT AND THE SURGERY COMPLETED WITHOUT FURTHER ISSUE. THE EVENT CAUSED NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8381969 ↗
  • MalfunctionFeb 21, 2019Break

    WHILE TAPPING THE PEDICLE, THE 5.5MM TAP BROKE. THE DETACHED SECTION WAS RETRIEVED FROM THE PATIENT AND THE SURGERY COMPLETED WITHOUT FURTHER ISSUE. THE EVENT CAUSED NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8357653 ↗

Other TAP, BONE devices

All TAP, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page