Skip to content

BLACKHAWK Cervical Spacer System S-TT10-1412C07Recalls & reported problems

Choice Spine, LP · BLACKHAWK TI,CON,STERILE,14WX12DX7HXC · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840996182201), received between Aug 6, 2025 and Aug 6, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Migration; Migration

    THE CAM OF THE BLACKHAWK TI IMPLANT DID NOT LOCK DURING THE INITIAL SURGERY. THE SURGEON NOTICED 4 WEEKS POST OP THAT ONE OF THE BLADES HAD BACKED OUT OF THE BONE. THE X RAY IMAGE INDICATED THAT A SCREW IN AN ADJACENT PLATE HAD PROHIBITED THE BLADES FROM BEING FULLY DEPLOYED AND NOT ALLOWING THE CAM TO LOCK APPROPRIATELY. ON (B)(6) 2025 A REVISION SURGERY WAS PERFORMED TO REMOVE THE SCREW FROM THE ADJACENT CERVICAL PLATE TO ENABLE THE BLADE ON THE BLACKHAWK TI IMPLANT TO FULLY DEPLOY AND ALLOW THE CAM TO LOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730127 ↗

Other Intervertebral fusion device with integrated fixation, cervical devices

All Intervertebral fusion device with integrated fixation, cervical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page