"1.5MM" SYSTEM 01-7144Recalls & reported problems
Biomet Microfixation, Inc · Class 2
All models of "1.5MM" SYSTEM (35)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Biomet MicroFixation 1.5mm system drill bit, used to drill bone so small screws can be placed. Most reports describe the drill bit fracturing or breaking during surgery, sometimes leaving fragments in the patient that had to be retrieved, with11 injuries and no deaths among 32 reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00841036013165), received between Apr 26, 2019 and Jul 7, 2025 — classified as 11 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Fracture” (29 reports).
1 other recall cover devices cleared under the same FDA submission (K062842); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture29
- Break2
- Difficult to Remove1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Foreign Body In Patient8
- No Known Impact Or Consequence To Patient7
- No Consequences Or Impact To Patient2
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Fracture; Fracture
IT IS REPORTED THAT WHILE THE SURGEON WAS USING THE DRILL BIT TO DRILL THROUGH THE MUCOSA AND BONE TO PLACE 1.5MM SCREWS FOR IMF FIXATION, THE DRILL BIT FRACTURED. THIS ALSO HAPPENED WITH ANOTHER DRILL BIT. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22412840 ↗MalfunctionJun 9, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN INITIAL PROCEDURE, A DRILL TIP FRACTURED. THE BROKEN TIP WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS COMPLETED USING BACKUP INSTRUMENTATION. THE SURGICAL TECHNIQUE WAS FOLLOWED. NO ADDITIONAL HARM TO THE PATIENT WAS REPORTED. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167410 ↗MalfunctionNov 5, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILL WAS BROKEN DURING A PROCEDURE. ALL DEBRIS/FRAGMENTS WERE RECOVERED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20612258 ↗InjuryAug 22, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 20028481 ↗MalfunctionAug 6, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19925341 ↗MalfunctionAug 6, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILL FRACTURED DURING USE. DURING THE SURGERY, THE NURSE REPORTED TO THE SURGEON THAT THE DRILL WAS FRACTURED. THE SURGEON REMOVED ALL OF THE PLATES AND RETRIEVED THE DRILL TIP. PATIENT WAS RE-FIXED WITH A NEW PLATE AND SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19925185 ↗MalfunctionAug 6, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILL FRACTURED DURING USE. THERE WERE NO CONSEQUENCES OR IMPACTS TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19919941 ↗MalfunctionApr 29, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19202966 ↗MalfunctionJul 17, 2023Fracture; Fracture
IT IS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE DRILL FRACTURED OFF. THE FRAGMENT WAS RECOVERED FROM THE PATIENT, AND THE PROCEDURE WAS COMPLETED USING OTHER DRILLS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17332724 ↗InjuryJun 19, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 17153612 ↗InjuryJun 13, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 17121267 ↗MalfunctionNov 22, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE THE TWIST DRILL FRACTURED. THE BROKEN DRILL TIP DID NOT REMAIN IN THE PATIENT'S BODY AND WAS RECOVERED. A DIFFERENT DRILL WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15842947 ↗MalfunctionJul 12, 2022Break
IT WAS REPORTED THAT THE SURGEON CHECKED THE TIP WHILE USING THE DRILL TO MAKE A PILOT HOLE AND FOUND IT TO BE BROKEN. HE CONFIRMED THE SURGICAL FIELD AND RECOVERED THE BROKEN DRILL TIP, BUT IT WAS LOST IN THE CLEANING PROCESS. THE SURGERY WAS COMPLETED USING ANOTHER DRILL. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14991274 ↗MalfunctionJun 28, 2022Fracture
IT WAS REPORTED THAT THE DRILL TIP FRACTURED DURING A ZYGOMATIC FRACTURE OPEN REDUCTION AND FIXATION. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14843142 ↗MalfunctionJun 16, 2022Fracture
IT WAS REPORTED THAT THE MORNING AFTER AN OPEN REDUCTION AND FIXATION PROCEDURE, THE DRILL WAS NOTED TO HAVE FRACTURED. HOWEVER, THE SURGEON DID NOT NOTICE THE FRACTURED DURING THE SURGERY. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14712860 ↗InjuryMay 9, 2022Break; Difficult to Remove
DRILL TIP BROKE AND WAS LEFT IN PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 14336736 ↗MalfunctionApr 26, 2022Fracture; Fracture
IT WAS REPORTED A MALE SURGEON BROKE THE TIP OF A DRILL WHILE DRILLING A PILOT HOLE DURING AN OPEN REDUCTION AND FIXATION OF THE MAXILLA. HE RECOVERED THE BROKEN TIP AND CONTINUED THE OPERATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14207838 ↗InjuryFeb 7, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A ZYGOMATIC FRACTURE OPEN BLOOD RESTORATION PROCEDURE, THE DRILL FRACTURED WHEN THE SURGEON ATTEMPTED TO MAKE THE FIRST HOLE IN THE PATIENT'S BONE. THE TIP OF THE DRILL REMAINS IN THE MAXILLARY SINUS. THERE IS NO SURGERY PLANNED TO REMOVE THE TIP. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 13472449 ↗MalfunctionSep 7, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12435252 ↗InjuryNov 30, 2020Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10917837 ↗InjuryOct 30, 2020Fracture; Fracture
IT WAS REPORTED THE DRILL TIP FRACTURED DURING A PROCEDURE. THE MAXILLARY ANTERIOR WALL WAS SECURED WITH A 1.5L PLATE AND THE DRILL BROKE DURING USE. THE NURSE NOTED THE FRACTURED PIECE COULD NOT BE LOCATED IN THE BODY AND THE DOCTOR TOOK AN X-RAY WHICH DID NOT SHOW THE FRACTURED PIECE. LATER THE FRACTURED PIECE OF THE DRILL WAS LOCATED IN THE SUCTION EFFLUENT BOTTLE AND THE PROCEDURE WAS RESUMED. THE SURGERY WAS COMPLETED WITH AN ALTERNATE DRILL. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10762730 ↗MalfunctionOct 1, 2020Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THE DRILL WAS DISCOVERED TO BE DAMAGED DURING A PROCEDURE ON THE INFRAORBITAL MARGIN. THE DRILL WAS INSPECTED PRIOR TO THE SURGERY, AND FOUND TO BE UNDAMAGED. PRIOR TO USE IN THE SURGERY, IT WAS FOUND TO BE DAMAGED, AND WAS NOT USED. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DRILL BIT. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10612917 ↗MalfunctionSep 15, 2020Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10538758 ↗MalfunctionJul 9, 2020Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10253373 ↗MalfunctionMar 16, 2020Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9836490 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062842. One submission often covers many models, so check the product and lot details in each record.
- Z-0538-2010 · Nov 6, 2009 · TerminatedBiomet 1.1 X 50MM, 3.5MM STOP STAINLESS STEEL, 1.5MM SYSTEM TWIST DRILL WITH J NOTCH, REF 01-7141. DISTRIBUTED BY: BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 32218
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page