"1.5MM" SYSTEM TWIST DRILL 01-7149Recalls & reported problems
Biomet Microfixation, Inc · Class 2
All models of "1.5MM" SYSTEM TWIST DRILL (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00841036013202), received between Aug 20, 2019 and Jul 18, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Fracture” (14 reports).
1 other recall cover devices cleared under the same FDA submission (K062842); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture14
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Known Impact Or Consequence To Patient4
- No Information1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2025Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF THE DRILL WAS FRACTURED AND RETRIEVED DURING A PROCEDURE. ANOTHER DRILL WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22541522 ↗MalfunctionAug 27, 2024Fracture; Fracture
IT IS REPORTED THAT THE DRILL FRACTURED DURING A PROCEDURE. ALL FRAGMENTS WERE RETRIEVED AND NO HARM OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20076136 ↗MalfunctionOct 3, 2022Fracture; Fracture
IT WAS REPORTED THAT THE DRILL FRACTURED WHEN TESTING THE ROTATION. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15528061 ↗MalfunctionSep 15, 2022Fracture; Fracture
IT WAS REPORTED THAT TWO DRILLS FRACTURED DURING DRILLING. THE FRACTURED PARTS OF BOTH DRILLS WERE EMBEDDED IN THE BONE, THE BONE AROUND THE EMBEDDED PARTS WAS SCRAPED OFF, AND ALL THE FRACTURED PARTS WERE REMOVED. THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS. THERE WERE NO ADVERSE EVENTS AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15434376 ↗MalfunctionDec 14, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13005530 ↗MalfunctionAug 31, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12391257 ↗MalfunctionJul 27, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12227667 ↗MalfunctionJun 22, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12044085 ↗MalfunctionMay 24, 2021Fracture; Fracture
IT WAS REPORTED THE TIP OF THE DRILL WAS MISSING AND IT WAS DISCOVERED UPON RECEIVING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11869600 ↗MalfunctionJun 29, 2020Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10206383 ↗MalfunctionApr 20, 2020Fracture; Fracture
IT WAS REPORTED THAT TWO DRILL BITS FRACTURED DURING SURGERY. THE SURGEON REPORTED THAT THE FRACTURE OCCURRED AT THE MOMENT OF CONTACT WITH THE BONE, AND THAT THE PATIENT HAD UNUSUALLY HARD BONE. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9977129 ↗MalfunctionDec 26, 2019Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Information
MAUDE report 9521140 ↗MalfunctionAug 30, 2019Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8952724 ↗MalfunctionAug 20, 2019Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8912138 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062842. One submission often covers many models, so check the product and lot details in each record.
- Z-0538-2010 · Nov 6, 2009 · TerminatedBiomet 1.1 X 50MM, 3.5MM STOP STAINLESS STEEL, 1.5MM SYSTEM TWIST DRILL WITH J NOTCH, REF 01-7141. DISTRIBUTED BY: BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 32218
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.