2.0MM SYSTEM 01-9196Recalls & reported problems
Biomet Microfixation, Inc · DRILL 1.5X50MM 22MM STP J-NT · Class 2
All models of 2.0MM SYSTEM (11)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00841036016968), received between Jan 17, 2020 and Apr 1, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Fracture” (11 reports).
1 other recall cover devices cleared under the same FDA submission (K062842); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture11
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2025Fracture; Fracture
IT WAS REPORTED THAT A DRILL FRACTURED NEAR THE ROOT DURING AN INITIAL FIXATION PROCEDURE. THE CORRECT SURGICAL TECHNIQUE WAS USED AND THE SURGEON NOTED THAT THE BREAKAGE MAY HAVE OCCURRED DUE TO THE APPLICATION OF OBLIQUE FORCE ON A LONG DRILL WITHOUT A GUIDE. THE FRACTURED DRILL TIP WAS SUCCESSFULLY REMOVED, AND THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21742672 ↗MalfunctionMar 28, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILL TIP FRACTURED WHILE IN USE DURING A PROCEDURE. THE DRILL TIP REMAINED IN THE PATIENT SO THE SURGEON PROCEEDED TO REMOVE THE FRAGMENT. THE SURGEON INJURED HIS FINGER WHILE REMOVING THE DRILL FRAGMENT. THERE WAS NO FURTHER INFORMATION PROVIDED FOR THE SURGEON'S INJURY. NO FRAGMENTS WERE RETAINED BY THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19001294 ↗MalfunctionFeb 1, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILLS FRACTURED WHILE DRILLING A HOLE FOR SCREW INSERTIONS DURING AN OSTEOTOMY PROCEDURE FOR JAW DEFORMITY. ANOTHER DRILL WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18627508 ↗MalfunctionFeb 1, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILLS FRACTURED WHILE DRILLING A HOLE FOR SCREW INSERTIONS DURING AN OSTEOTOMY PROCEDURE FOR JAW DEFORMITY. ANOTHER DRILL WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18627499 ↗MalfunctionDec 18, 2023Fracture; Fracture
IT IS REPORTED THAT DURING AN OSTEOTOMY PROCEDURE, THE DRILL TIP FRACTURED WHILE DRILLING TO INSERT A SCREW. NO FRAGMENTS WERE RETAINED BY THE PATIENT, AND THERE WAS NO HARM OR INJURY REPORTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DRILL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18345479 ↗MalfunctionNov 7, 2023Fracture; Fracture
IT WAS REPORTED THAT THE DRILL FRACTURED DURING AN INTERMAXILLARY FIXATION PROCEDURE. ANOTHER DRILL WAS USED TO COMPLETE THE PROCEDURE.THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18084769 ↗MalfunctionNov 7, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18084745 ↗MalfunctionAug 9, 2022Fracture; Fracture
IT WAS REPORTED THE DRILL TIP FRACTURED WHEN THE SURGEON FIXED THE PLATE TO THE MANDIBLE. THE DEVICE WAS NOT LEFT IN THE PATIENT AND THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15192245 ↗MalfunctionNov 30, 2020Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10917768 ↗MalfunctionFeb 19, 2020Fracture; Fracture
IT WAS REPORTED THAT A DRILL BIT FRACTURED DURING SURGERY. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9730206 ↗MalfunctionJan 17, 2020Fracture; Fracture
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9603688 ↗
FDA clearance and classification
Cleared through 510(k) K062842 on Dec 4, 2006 (decision: Substantially Equivalent). Submitted by Walter Lorenz Surgical, Inc. as “Twist Drill”. The FDA review took 73 days. FDA record ↗
The FDA regulates product code HBE (drills, burrs, trephines & accessories (simple, powered)) under 21 CFR 882.4310 as a Class II device. 21 CFR 882.4310 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K062842. One submission often covers many models, so check the product and lot details in each record.
- Z-0538-2010 · Nov 6, 2009 · TerminatedBiomet 1.1 X 50MM, 3.5MM STOP STAINLESS STEEL, 1.5MM SYSTEM TWIST DRILL WITH J NOTCH, REF 01-7141. DISTRIBUTED BY: BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 322
Common questions about the 2.0MM SYSTEM 01-9196
Has the 2.0MM SYSTEM 01-9196 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00841036016968), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the 2.0MM SYSTEM 01-9196?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Jan 17, 2020 and Apr 1, 2025. In those reports, the most common specific device problems are “Fracture” (11) and “Material Twisted/Bent” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the 2.0MM SYSTEM 01-9196?
None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the 2.0MM SYSTEM 01-9196 FDA approved?
It was cleared through 510(k) K062842 on Dec 4, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the 2.0MM SYSTEM 01-9196, and is it still sold?
The FDA's device database lists Biomet Microfixation, Inc as the company responsible for it (brand 2.0MM SYSTEM, model 01-9196). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Drills, Burrs, Trephines & Accessories (Simple, Powered) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,624 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)930 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)414 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)221 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)186 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)184 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)182 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)164 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page