Skip to content

PediLoc Tibial Plate, LCB 00-0903-2655Recalls & reported problems

Orthopediatrics Corp. · 3.5MM LOCKING CORTICAL SCREW WITH T15 HEXALOBE, SIZE 55MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132111055), received between Dec 18, 2025 and Dec 18, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 18, 2025Material Twisted/Bent; Material Twisted/Bent

    THREE DAYS POST-OPERATION, THE PATIENT MOVED, AND FOLLOW-UP X-RAYS REVEALED SIGNIFICANT BACKING OUT OF THE Z-HOLE LOCKING SCREW, WITH THE PROXIMAL FEMUR/HEAD HAVING DISPLACED FROM THE DISTAL FEMUR. A REVISION PROCEDURE WAS REQUIRED TO RESTORE ALIGNMENT BETWEEN THE PROXIMAL AND DISTAL FEMUR. DURING THE REVISION, THE EXISTING PROXIMAL SCREWS WERE REPLACED, AND IT WAS NOTED THAT THE REMOVED SCREWS WERE SUBSTANTIALLY BENT.

    Patient problems: Failure of Implant

    MAUDE report 23837947 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page