Skip to content

PediLoc Fragment System 00-1005-4250Recalls & reported problems

Orthopediatrics Corp. · 4.0mm Cannulated Cancellous Screw, partial thread, T15, 50mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132113493), received between Feb 20, 2018 and Feb 20, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Material Disintegration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Disintegration1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2018Material Disintegration; Material Disintegration

    IT HAS BEEN REPORTED THAT DURING AN IMPLANTATION OF A PEDIFRAG PLATE, A SCREW MET A PREVIOUSLY INSERTED CORTICAL SCREW CAUSING THE SCREW TO UNRAVEL IN THE PATIENT. THE DAMAGED SCREW WAS REMOVED AND A NEW SCREW INSERTED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7284059 ↗

Other Screw, fixation, bone devices