Type-S™ Extension MTCFHNVRecalls & reported problems
Medtec, Inc. · For use with Varian · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841439104156), received between May 7, 2020 and Jun 24, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
1 FDA recall record name this device's identifier.
1 other recall cover devices cleared under the same FDA submission (K933227); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 12, 2020Terminated
Potential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.
Z-2243-2020 · Root cause: Component design/selection · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component2
Patient problems
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2020Detachment of Device or Device Component
EXTENSION DISENGAGED FROM THE TABLE WHEN TWO THERAPISTS LIFTED THE PATIENT ON A SHEET WHILE THE PATIENT WAS UNDER A THERMOPLASTIC MASK. PATIENT WAS SUPPORTED BY THERAPISTS HOLDING SHEET AND DID NOT FALL. AFTER TAKING KV IMAGES, IT WAS DECIDED THAT THE PATIENT NEEDED ADJUSTMENT FOR PROPER ALIGNMENT FOR TREATMENT. THERAPISTS USED A SHEET UNDERNEATH THE PATIENT TO MOVE THE PATIENT SLIGHTLY FURTHER UP INTO THE HEAD, NECK, AND SHOULDER THERMOPLASTIC MASK AND WHILE DOING SO, THE EXTENSION DETACHED FROM THE COUCH. THE PATIENT STARTED TO FALL WITH THE EXTENSION BUT WAS CAUGHT BY THE THERAPISTS. THERE WERE NO INJURIES; THE PATIENT WAS JUST ALARMED AND SCARED DURING THE INCIDENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10189608 ↗InjuryMay 7, 2020Detachment of Device or Device Component
EXTENSION DISENGAGED FROM THE TABLE WHEN TWO THERAPISTS LIFTED THE PATIENT ON A SHEET WHILE THE PATIENT WAS UNDER A THERMOPLASTIC MASK. PATIENT WAS SUPPORTED BY THERAPISTS HOLDING SHEET AND DID NOT FALL. AFTER TAKING KV IMAGES, IT WAS DECIDED THAT THE PATIENT NEEDED ADJUSTMENT FOR PROPER ALIGNMENT FOR TREATMENT. THERAPISTS USED A SHEET UNDERNEATH THE PATIENT TO MOVE THE PATIENT SLIGHTLY FURTHER UP INTO THE HEAD, NECK, AND SHOULDER THERMOPLASTIC MASK AND WHILE DOING SO, THE EXTENSION DETACHED FROM THE COUCH. THE PATIENT STARTED TO FALL WITH THE EXTENSION BUT WAS CAUGHT BY THE THERAPISTS. THERE WERE NO INJURIES; THE PATIENT WAS JUST ALARMED AND SCARED DURING THE INCIDENT.
Patient problems: No Code Available
MAUDE report 10034120 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K933227. One submission often covers many models, so check the product and lot details in each record.
- Z-0124-2010 · Aug 31, 2009 · TerminatedCIVCO MTAPSID2932 Type-S Thermoplastic, Disposable nonsterile (48.3 cm) IMRT Style 29 thermoplastic (3.2 mm), Ref. MTAPSID2932, disposable frame mask, 5 per box. The firm name on
Other Accelerator, linear, medical devices
- LANDAUER microSTAR® ii Dosimetry SystemLandauer, Inc. · Accelerator, Linear, Medical38 recalls0 reports
- MOSAIQElekta, Inc. · Accelerator, Linear, Medical3 recalls112 reports
- Elekta Medical Linear Accelerator 1552093Elekta Solutions AB · Accelerator, linear, medical5 recalls6 reports
- Elekta Medical Linear Accelerator 1552092Elekta Solutions AB · Accelerator, linear, medical5 recalls0 reports
- MOSAIQ 3.2Elekta Solutions AB · Accelerator, Linear, Medical4 recalls25 reports
- Elekta Medical Linear Accelerator 1550358Elekta Solutions AB · Accelerator, linear, medical4 recalls2 reports
- Elekta Medical Linear Accelerator 1552091Elekta Solutions AB · Accelerator, linear, medical4 recalls1 report
- Elekta Medical Linear Accelerator 1550359Elekta Solutions AB · Accelerator, linear, medical4 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.