Skip to content

APEX Knee System KR-46000Recalls & reported problems

Omni Life Science, Inc. · Modular Tibia Augment Bolt Driver · Class 1

All models of APEX Knee System (30)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841690117841), received between Nov 14, 2024 and Nov 14, 2024 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE SURGERY ON (B)(6) 2024. DOCTOR WAS DRILLING THE TIBIA CANAL WITH A BOSS REAMER. IT WAS IDENTIFIED BY THE REP ATTENDING THE SURGERY TO THE SURGEON THAT A SMALL PART OF THE REAMER WAS MISSING. THE SCRUB NURSE SEARCHED FOR THE DEVICE AND COULD NOT FIND IT. THE SURGEON STATED THAT IF IT IS IN THE TIBIAL HE IS NOT PREPARED TO REMOVE THE TIBIAL REVISION IMPLANT AS IT WOULD BE MORE DETRIMENTAL TO THE PATIENT. THAT THEY WOULD X-RAY POST OPERATIVELY IF NOT FOUND. SUBSEQUENT TO SURGERY A POST OP X-RAY WAS PERFORMED IN RECOVERY ROOM IDENTIFYING THAT THE INSTRUMENT IS IN THE TIBIAL SHAFT - DISTAL TO THE TIBIAL STEM AND NOT IN CONTACT AS SURROUNDED BY PALACOS R ANTIBIOTIC CEMENT. THIS HAD BEEN IDENTIFIED TO THE SCRUB AT THE START OF THE CASE, WHICH IS WHY A COUNT OF THIS ITEM COMMENCED PRIOR TO CLOSING THE SURGICAL WOUND. ALLOWING THE SURGEON TO INTERVENE IF HE DEEMED APPROPRIATE.

    Patient problems: Foreign Body In Patient

    MAUDE report 20692307 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page