Skip to content

Hintermann Series H2 HT-32802Recalls & reported problems

Vilex LLC · Hintermann Series H2 Tibia Implant, Sz2 Left, 7mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841731142139), received between Mar 30, 2026 and Mar 30, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Defective Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Defective Component

    VILEX BECAME AWARE OF AN EVENT INVOLVING THE H2 TOTAL ANKLE IMPLANT COMPONENTS ON (B)(6) 2026. THE PURPOSE OF THE EVENT SURGERY WAS TO REVISE AN EXISTING TOTAL ANKLE REPLACEMENT SYSTEM FROM A DIFFERENT MANUFACTURER. REP STATES THE TIBIAL SLIDE LOCKING MECHANISM, PACKAGED WITH THE TIBIAL COMPONENT, WAS INSERTED IN THE CORRECT ORIENTATION AND FAILED TO LOCK THE INSERTED POLY COMPONENT IMPLANT. ONCE IT WAS DETERMINED THAT THE LOCK WOULD NOT ENGAGE WITH THE POLY COMPONENT, THE TIBIAL AND POLY COMPONENTS WERE REMOVED, AND A NEW TIBIAL COMPONENT ASSEMBLY AND POLY ASSEMBLY WERE INSERTED. TOTAL DELAY IN SURGERY APPROXIMATELY 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715758 ↗
  • MalfunctionMar 30, 2026Defective Component

    DURING INVESTIGATION OF REPORTED COMPLAINT (B)(4) (REPORT NO. 1051526-2026-00003), THE FOLLOWING ADDITIONAL EVENT WAS IDENTIFIED. DURING AN H2 TOTAL ANKLE SURGERY, PERFORMED (B)(6) 2025, THE TIBIAL SLIDE LOCKING MECHANISM THAT WAS PACKAGED WITH THE TIBIAL IMPLANT FAILED TO LOCK THE POLY INLAY COMPONENT IN PLACE. THE LOCKING MECHANISM WAS REMOVED AND REPLACED WITH A NEW MECHANISM THAT SUCCESSFULLY LOCKED THE POLY INLAY INTO PLACE AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions