Preat 9000662Recalls & reported problems
Preat Corporation · NobelBiocare™ Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base Kit contains 1 NobelBiocare™ Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base 2 NobelBiocare™ Tri-Lobe-compatible RP/5.0mm/6.0mm Titanium Implant Screws · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K183518); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jun 25, 2025Class IIOpen, Classified
Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.
Z-2514-2025 · Root cause: Process control · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K183518 on Mar 18, 2019 (decision: Substantially Equivalent). Submitted by Preat Corporation as “Preat Abutments”. The FDA review took 90 days. FDA record ↗
The FDA regulates product code NHA (abutment, implant, dental, endosseous) under 21 CFR 872.3630 as a Class II device. 21 CFR 872.3630 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: To be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.
Recalls under the same FDA submission
These recalls involve devices cleared under K183518. One submission often covers many models, so check the product and lot details in each record.
- Z-1987-2025 · Apr 2, 2025 · Open, ClassifiedBiomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary
- Z-1321-2024 · Feb 9, 2024 · Open, ClassifiedPreat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967
Common questions about the Preat 9000662
Has the Preat 9000662 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2514-2025 (Class II), began on Jun 25, 2025: Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the Preat 9000662 FDA approved?
It was cleared through 510(k) K183518 on Mar 18, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Preat 9000662 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Preat 9000662, and is it still sold?
The FDA's device database lists Preat Corporation as the company responsible for it (brand Preat, model 9000662). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Abutment, implant, dental, endosseous devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS150 reports
- Nobel Biocare® 36662Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS2 recalls5 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous61 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- CERTAIN® BELLATEK® ILDAN3ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page