Skip to content

Fusion Cups FC10Recalls & reported problems

Trimed, INC. · Fusion Cup, 10 Hole, 18mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842188100734), received between Dec 15, 2022 and Dec 22, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Crack1
  • Fracture1
  • Material Integrity Problem1

Patient problems

  • Implant Pain2
  • Arthritis1
  • Rheumatoid Arthritis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 22, 2025Crack; Material Integrity Problem; Crack; Material Integrity Problem

    SALES REP REPORTED ON 26NOV2025 THAT A FC10 HAD TO BE REMOVED FROM A PATIENT DUE TO A NON-UNION. AFTER THE DEVICE WAS REMOVED, IT WAS DISCOVERED THAT THE FC10 WAS CRACKED/BROKEN. THE DEVICE AND THE REPS' FIELD FEEDBACK FORM WAS MAILED TO TRIMED AND RECEIVED ON 11DEC2025 WHICH IS WHEN THE INFORMATION ABOUT THE RETURN WAS FORWARDED TO THE QUALITY DEPARTMENT. QUALITY FOLLOWED UP WITH THE REP ON 11DEC2025 WHICH THE REP PROVIDED ADDITIONAL INFORMATION ON 15DEC2025. THE REP STATED THAT THEY ARE NOT SURE WHEN THE DEVICE WAS ORIGINALLY IMPLANTED, HE ASSUMES THE DISCOMFORT WAS RELATED TO THE FACT THAT THIS WAS A NON-UNION, HE DOES NOT KNOW WHAT WAS USED TO REPLACE THE DEVICE AND HE DOES NOT KNOW WHAT CAUSED THE NON-UNION. PATIENT RELATED INFORMATION IS UNKNOWN. THE REMOVAL SURGERY WAS SUCCESSFUL BUT THERE WERE NO X-RAYS PROVIDED.

    Patient problems: Implant Pain

    MAUDE report 23865591 ↗
  • InjuryJun 14, 2023Fracture; Fracture

    THE SURGEON STATE THAT THE POSITION OF THE PLATE WAS NOT IDEAL THAT COULD HAVE CAUSED THE DELAYED UNION WHICH WOULD HAVE LEAD TO FATIGUE OF THE PLATE. THE SURGEON STATED THAT IT WAS BELIEVED THAT THE POSITION OF THE PLATE WAS THE PRIMARY PROBLEM OF THE PATIENT DISCOMFORT, RATHER THAN MATERIAL FAILURE.

    Patient problems: Implant Pain

    MAUDE report 17131897 ↗
  • MalfunctionDec 15, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    (B)(6) REPORTED ON (B)(6)2022 THAT HE WOULD LIKE TO RETURN A FC10 MARKED WITH "DO NOT IMPLANT" ON IT. IT WAS REPORTED THAT THIS ITEM DID NOT COME IN CONTACT WITH A PATIENT AND DID NOT CAUSE ADDITIONAL HARM. THE ITEM IS AVAILABLE TO BE RETURNED. PICTURES ARE AVAILABLE.

    Patient problems: Arthritis; Rheumatoid Arthritis

    MAUDE report 15993139 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates