Skip to content

Osteotomy Compression Plates UOCP-SRecalls & reported problems

Trimed, INC. · Ulnar Osteotomy Comp Plate Short · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842188101533), received between Mar 22, 2018 and Mar 22, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K043263); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 22, 2018Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2018, TRIMED WAS NOTIFIED THAT TWO HOLES, OF UOCP-S PLATE, THAT TAKE THE SAW GUIDE WERE ON THE WRONG SIDE OF THE PLATE. THE SURGEON WENT TO LEAVE THE PLATE IN-SITU, AS THE ISSUE ONLY CAME TO LIGHT ONCE THE OSTEOTOMY WAS PERFORMED. THE SURGEON USED A LAG SCREW THROUGH THE SLOTTED HOLE INSTEAD, THE OSTEOTOMY WAS STABLE. TRIMED WAS NOTIFIED BY A DISTRIBUTOR, (B)(4), IN (B)(6). THIS EVENT HAPPENED IN A CLINICAL SETTING, AND THIS IS THE FIRST EVENT OF ITS KIND FOR THE UOCP-S PLATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7363139 ↗

FDA clearance and classification

Cleared through 510(k) K043263 on Jan 14, 2005 (decision: Substantially Equivalent). Submitted by TriMed, Inc. as “Trimed Osteotomy Plate”. The FDA review took 51 days. FDA record ↗

The FDA regulates product code HRS (plate, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K043263. One submission often covers many models, so check the product and lot details in each record.

  • Z-2877-2018 · Apr 16, 2018 · Terminated
    Ulnar Osteotomy Compression Plate, Short Model: UOCP-S Product Usage: The UOCP-S is provided non-sterile with the required fixation screws and surgical instrumentation. The TriMed

Common questions about the Osteotomy Compression Plates UOCP-S

Has the Osteotomy Compression Plates UOCP-S been recalled?

No FDA recall record in our data names this exact device identifier (DI 00842188101533), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Osteotomy Compression Plates UOCP-S?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 22, 2018. That report lists device problem “Manufacturing, Packaging or Shipping Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Osteotomy Compression Plates UOCP-S?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Osteotomy Compression Plates UOCP-S FDA approved?

It was cleared through 510(k) K043263 on Jan 14, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Osteotomy Compression Plates UOCP-S MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Osteotomy Compression Plates UOCP-S, and is it still sold?

The FDA's device database lists Trimed, INC. as the company responsible for it (brand Osteotomy Compression Plates, model UOCP-S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates