Skip to content

Tools - Screws PEG-XTNDRRecalls & reported problems

Trimed, INC. · Peg Guide Extender · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00842188112867), received between Dec 20, 2017 and Feb 9, 2018 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Corroded” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Corroded4
  • Material Integrity Problem2
  • Device Inoperable1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Patient Involvement2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2018Material Integrity Problem; Material Integrity Problem

    ON (B)(6) 2018, IT WAS REPORTED, BY A TRIMED SALES REP, THAT A PEG-XTNDR IN HIS SET WERE SHOWING RUST AROUND THE SPRING. THE REP REPORTED THAT THIS FEEDBACK WAS NOT ASSOCIATED WITH A CLINICAL EVENT, AND THERE WAS NO PATIENT INVOLVEMENT. THIS IS THE SIXTH COMPLAINT OF THIS KIND FOR THIS TYPE OF PART.

    Patient problems: No Patient Involvement

    MAUDE report 7261847 ↗
  • MalfunctionFeb 9, 2018Material Integrity Problem; Material Integrity Problem

    ON JANUARY 11, 2018, IT WAS REPORTED, BY A TRIMED SALES REP, THAT A PEG-XTNDR IN HIS SET WERE SHOWING RUST AROUND THE SPRING. THE REP REPORTED THAT THIS FEEDBACK WAS NOT ASSOCIATED WITH A CLINICAL EVENT, AND THERE WAS NO PATIENT INVOLVEMENT. THIS IS THE SIXTH COMPLAINT OF THIS KIND FOR THIS TYPE OF PART.

    Patient problems: No Patient Involvement

    MAUDE report 7261846 ↗
  • MalfunctionDec 20, 2017Corroded; Device Inoperable

    DURING VISUAL REVIEW OF MATERIAL BEING SENT TO HOSPITAL, IT WAS DISCOVERED THAT THE SPRING ON A PEG-XTNDR HAD CORRODED. THE NON-CONFORMING PART WAS RETURNED TO THE DISTRIBUTOR, AT WHICH POINT THE DISTRIBUTOR NOTIFIED TRIMED OF THE COMPLAINT ON (B)(6) 2017. THIS INCIDENT OCCURRED IN CHILE. THIS TYPE OF MALFUNCTION FOR THIS PART HAS AN OCCURRENCE RATE OF (B)(4) % BASED ON SALES SINCE 2010.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7136685 ↗
  • MalfunctionDec 20, 2017Corroded

    DURING VISUAL REVIEW OF MATERIAL BEING SENT TO HOSPITAL, IT WAS DISCOVERED THAT THE SPRING ON A PEG-XTNDR HAD CORRODED. THE NON-CONFORMING PART WAS RETURNED TO THE DISTRIBUTOR, AT WHICH POINT THE DISTRIBUTOR NOTIFIED TRIMED OF THE COMPLAINT ON (B)(6) 2017. THIS INCIDENT OCCURRED IN CHILE. THIS TYPE OF MALFUNCTION FOR THIS PART HAS AN OCCURRENCE RATE OF (B)(4) BASED ON SALES SINCE 2010.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7136682 ↗
  • MalfunctionDec 20, 2017Corroded

    DURING VISUAL REVIEW OF MATERIAL BEING SENT TO HOSPITAL, IT WAS DISCOVERED THAT THE SPRING ON A PEG-XTNDR HAD CORRODED. THE NON-CONFORMING PART WAS RETURNED TO THE DISTRIBUTOR, AT WHICH POINT THE DISTRIBUTOR NOTIFIED TRIMED OF THE COMPLAINT ON (B)(6) 2017. THIS INCIDENT OCCURRED IN CHILE. THIS TYPE OF MALFUNCTION FOR THIS PART HAS AN OCCURRENCE RATE OF 0.24% BASED ON SALES SINCE 2010.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7136680 ↗
  • MalfunctionDec 20, 2017Corroded

    DURING VISUAL REVIEW OF MATERIAL BEING SENT TO HOSPITAL, IT WAS DISCOVERED THAT THE SPRING ON A PEG-XTNDR HAD CORRODED. THE NON-CONFORMING PART WAS RETURNED TO THE DISTRIBUTOR, AT WHICH POINT THE DISTRIBUTOR NOTIFIED TRIMED OF THE COMPLAINT ON (B)(6) 2017. THIS INCIDENT OCCURRED IN (B)(6). THIS TYPE OF MALFUNCTION FOR THIS PART HAS AN OCCURRENCE RATE OF (B)(4) BASED ON SALES SINCE 2010.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7136679 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.