Skip to content

WEB Aneurysm Embolization System W4-7-4-MVI-3Recalls & reported problems

Microvention, Inc. · WEB 21 SL 7x4 · Class 3 · Implant

All models of WEB Aneurysm Embolization System (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842429107096), received between Mar 17, 2021 and Jul 12, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Separation Failure1

Patient problems

  • Confusion/ Disorientation1
  • Hematoma1
  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1
  • Stroke/CVA1
  • Thrombosis/Thrombus1
  • Vomiting1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 12, 2024Separation Failure; Separation Failure

    IT WAS REPORTED THAT THE WEB DEVICE WAS POSITIONED BUT WOULD NOT DETACH WITH MULTIPLE DETACHMENT ATTEMPTS WITH THE DETACHMENT CONTROLLER. THE DEVICE WAS WITHDRAWN FROM THE PATIENT AND THE CASE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THE PATIENT WAS REPORTED TO BE NORMAL AND THERE WAS NO REPORTED PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19727621 ↗
  • InjuryMar 17, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AN UNRUPTURED ANTERIOR COMMUNICATING ARTERY (ACA) ANEURYSM WAS TREATED WITH A WEB DEVICE. THE DEVICE WAS IMPLANTED WITH NO ISSUES, AND CONTROL ANGIOGRAMS SHOWED NO FLOW DELAY IN THE EITHER ACA. AN ANTERIOR FRONTAL M3 BRANCH WAS THEN FOUND TO BE OCCLUDED. INTRA-ARTERIAL TPA 4MG WAS ADMINISTERED AND A STENT RETRIEVER WERE USED, AND RECANALIZATION WAS ACHIEVED. THE PATIENT HAD NOT BEEN HEPARINIZED FOR 20 MINUTES DURING IMPLANTATION OF THE DEVICE AS THE ACT MACHINE WAS BROKEN AND A NEW ONE HAD TO BE RETRIEVED. 7 HOURS POST-PROCEDURE, THE PATIENT VOMITED AND WAS CONFUSED. A CT SHOWED A LEFT FRONTAL HEMATOMA, CAUSED BY THE IA TPA AND MICROWIRE MANIPULATION, AND ALSO INDICATION OF BILATERAL DISTAL ACA STROKE. IT IS THE PHYSICIAN'S OPINION THAT THE CLOT IN THE M3 WAS MOST LIKELY NOT ASSOCIATED WITH THE WEB DEVICE, BUT THE THAT THE BILATERAL ACA STROKES WERE MOST LIKELY CAUSED BY MICRO-EMBOLI FROM THE PROXIMAL SURFACE OF THE DEVICE. IT WAS RELATED THAT THE PATIENT IS NOW IN A LONG TERM ACUTE CARE FACILITY AND HAS "RESUMED MODIFIED ORAL DIET AND HAS STARTED SAYING A FEW WORDS. STILL IS DEPENDENT ON NURSING CARE FOR ACTIVITIES OF DAILY LIVING."

    Patient problems: Confusion/ Disorientation; Hematoma; Obstruction/Occlusion; Stroke/CVA; Thrombosis/Thrombus; Vomiting

    MAUDE report 11509913 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0170-2023 · Sep 21, 2022 · Open, Classified
    WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Other Intrasaccular Flow Disruption Device devices

All Intrasaccular Flow Disruption Device devices →

Sources and updates