Skip to content

WEB Aneurysm Embolization System W2-9-5-MVI-3Recalls & reported problems

Microvention, Inc. · WEB 27 SL 9x5 · Class 3 · Implant

All models of WEB Aneurysm Embolization System (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00842429107201), received between May 16, 2023 and Sep 18, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Difficult or Delayed Separation1
  • Material Invagination1

Patient problems

  • Aneurysm2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED THROUGH A WEB PAS CLINICAL STUDY, AS REPORTED THAT AT 1 YEAR VISIT CEREBRAL ANGIOGRAM SHOWED COMPACTION AND RESIDUAL NECK REMNANT AT THE TREATED ANEURYSM. THE AE OUTCOME IS ONGOING. THE PATIENT IS ASYMPTOMATIC AND WILL REQUIRE RETREATMENT.

    Patient problems: Aneurysm

    MAUDE report 20252878 ↗
  • InjuryMar 18, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THROUGH THE WEB CLINICAL TRIAL THAT A WEB DEVICE USED TO TREAT AN ANEURYSM MIGRATED CRANIALLY AND TO THE LEFT, WITH A RECURRENT ANEURYSM FORMATION AT THE NECK OF THE TREATED ANEURYSM. IT WAS REPORTED THAT RETREATMENT WAS DEEMED NECESSARY IN THE NEAR FUTURE. THE PATIENT WAS ALSO REPORTED TO HAVE EXPERIENCED SEVERE HEADACHES.

    Patient problems: Aneurysm

    MAUDE report 18926731 ↗
  • MalfunctionJun 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION RECEIVED NOTES THE WDC-2 WAS NEVER USED ON THE WEB DEVICE ONCE IT ENTERED THE PATIENT. IT WAS ONLY USED DURING THE PREP TO CHECK THE WEB DEVICE. THE PHYSICIAN¿S TREATMENT PLAN WAS FOR THE FREDX 27 TO OPEN AND COVER THE ANEURYSM. BECAUSE THE FREDX 27 DID NOT OPEN HE DECIDED TO USE THE PIPELINE FLOW DIVERTER. THE HOSPITAL DOES NOT KEEP PIPELINE FLOW DIVERTERS IN THEIR INVENTORY, SO THE PATIENT NEEDED TO BE RESCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17040605 ↗
  • MalfunctionMay 16, 2023Material Invagination; Difficult or Delayed Separation; Material Invagination; Difficult or Delayed Separation

    SEE H10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16943270 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0170-2023 · Sep 21, 2022 · Open, Classified
    WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Other Intrasaccular Flow Disruption Device devices

All Intrasaccular Flow Disruption Device devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page