Skip to content

WEB W5-6-S-MVI-3Recalls & reported problems

Microvention, Inc. · WEB 17 SLS 6 · Class 3 · Implant

All models of WEB (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842429110911), received between Jul 16, 2025 and Jul 16, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Separation Failure; Separation Failure

    IT WAS REPORTED THAT VERY EXPERIENCED PRACTITIONER WHO IS FAMILIAR WITH THE WEB DEVICE WAS PLACING THIS DEVICE IN THE PATIENT'S ACOM VIA THE RIGHT SIDED, ARTERIAL APPROACH. DEVICE WAS SUCCESSFULLY DEPLOYED THROUGH A CATHETER AND LOOKED PERFECT. THE DEVICE WOULD NOT DETACH AND TRIED MULTIPLE DETACHMENT'S, TWO DETACHERS. AN ATTEMPT TO BRING THE CATHETER CLOSE TO THE DEVICE ONLY PROVED INEFFECTIVE AND RECAPTURED THE DEVICE. DEVICE WAS SUCCESSFULLY REMOVED, AND A DIFFERENT SIZE WEB WAS THEN USED TO SUCCESSFULLY TREAT THE PATIENT. NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22527697 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0170-2023 · Sep 21, 2022 · Open, Classified
    WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Other Intrasaccular Flow Disruption Device devices

All Intrasaccular Flow Disruption Device devices →