WEB W5-6-S-MVI-3Recalls & reported problems
Microvention, Inc. · WEB 17 SLS 6 · Class 3 · Implant
All models of WEB (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842429110911), received between Jul 16, 2025 and Jul 16, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2025Separation Failure; Separation Failure
IT WAS REPORTED THAT VERY EXPERIENCED PRACTITIONER WHO IS FAMILIAR WITH THE WEB DEVICE WAS PLACING THIS DEVICE IN THE PATIENT'S ACOM VIA THE RIGHT SIDED, ARTERIAL APPROACH. DEVICE WAS SUCCESSFULLY DEPLOYED THROUGH A CATHETER AND LOOKED PERFECT. THE DEVICE WOULD NOT DETACH AND TRIED MULTIPLE DETACHMENT'S, TWO DETACHERS. AN ATTEMPT TO BRING THE CATHETER CLOSE TO THE DEVICE ONLY PROVED INEFFECTIVE AND RECAPTURED THE DEVICE. DEVICE WAS SUCCESSFULLY REMOVED, AND A DIFFERENT SIZE WEB WAS THEN USED TO SUCCESSFULLY TREAT THE PATIENT. NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22527697 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.
- Z-0170-2023 · Sep 21, 2022 · Open, ClassifiedWEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Other Intrasaccular Flow Disruption Device devices
- WEB Detachment Controller WDC-2-MVIMicrovention, Inc. · Intrasaccular Flow Disruption Device1 recall0 reports
- WEB W5-4-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death23 reports
- WEB 17 SL 6x2 W5-6-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death20 reports
- WEB W5-5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device20 reports
- WEB W5-4.5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device19 reports
- WEB W5-7-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death16 reports
- WEB W5-6-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death15 reports
- WEB W5-5-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page