WEB 17 SL 7x2 W5-7-2-MVI-3Recalls & reported problems
Microvention, Inc. · WEB Aneurysm Embolization System · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00842429118009), received between Oct 30, 2023 and Oct 18, 2025 — classified as 1 death, 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure3
- Adverse Event Without Identified Device or Use Problem2
- Difficult or Delayed Separation2
- Detachment of Device or Device Component1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Cerebral Edema1
- Extravasation1
- Intracranial Hemorrhage1
- Intracranial Pressure Increased1
- Ruptured Aneurysm1
- Thrombosis/Thrombus1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 18, 2025Separation Failure; Separation Failure
AS REPORTED: PREPPED THE DEVICE ON THE BACK TABLE, DEPLOYED DEVICE IN SALINE AND GOT A GREEN LIGHT FROM THE DETACHER. DEPLOYED WEB IN ACOMM ANEURYSM AND CONTINUED TO GET A RED LIGHT WHEN WE WERE READY TO DETACH. TRIED A SECOND DETACHER AND CONTINUED TO BLINK RED. REMOVED THE WEB AND DEPLOYED A SECOND WEB AND DETACHED WITHOUT EVENT. THE DEVICE WAS DISCARDED BY THE USER FACILITY. NO ADVERSE EVENTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23331504 ↗MalfunctionOct 17, 2024Separation Failure; Separation Failure
IT WAS REPORTED THAT THE PHYSICIAN APPROACHED A RUPTURED PCOM ANEURYSM WITH AN ENVOY 6F 90CM CATHETER, A SOFIA 115CM, A STRAIGHT VIA17 AND A SYNCHRO14 GUIDEWIRE. THE PHYSICIAN THEN CHOSE A WEB 7X2MM DUE TO THE ANEURYSM SIZE (MEAN WIDTH 5MM, INFERIOR HEIGHT 5MM) AND PROPERLY PREPARED AND POSITIONED THE DEVICE INSIDE THE ANEURYSM. ONCE THE FIRST DETACHMENT ATTEMPT WAS MADE WITH A PROPER LOCK TENSION CONTROLLER MANEUVER AND WITH THE RIGHT SOUND AND LIGHT RESPONSES FROM THE WEB DETACHMENT CONTROLLER, THE DEVICE REMAINED CONNECTED TO ITS PUSHERWIRE. THE PHYSICIAN TRIED THEN TO DETACH THROUGH SOME PUSH AND PULL MANEUVERS WITH THE PUSHERWIRE OF THE WEB, THE MICROCATETHER AND EVENTUALLY THE SOFIA CATHETER, BUT THE WEB REMAINED CONNECTED TO THE PUSHERWIRE. DURING THE LAST ATTEMPT, THE PHYSICIAN PULLED THE WEB BACK INTO THE VESSEL AND DECIDED TO RESHEATH IT INTO THE SOFIA CATHETER AND NOT INTO THE VIA17 TO AVOID UNINTENTIONAL DETACHMENT DUE TO THE VIA17 INNER LUMEN SIZE. THEY SUCCESSFULLY PULLED SOFIA, VIA17 AND WEB OUT OF THE PATIENT, WITHOUT SEQUELAE AS CONFIRMED BY THE DIAGNOSTIC SERIES PERFORMED. THEN THE PHYSICIAN DECIDED TO USE A WEB SL 6X3MM WITH ANOTHER SOFIA 115CM AND A VIA17 MICROC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20477086 ↗MalfunctionJun 3, 2024Detachment of Device or Device Component; Difficult or Delayed Separation; Detachment of Device or Device Component; Difficult or Delayed Separation
IT WAS REPORTED THAT THE WEB DEVICE EXPERIENCED A DELAYED DETACHMENT. WHEN RETRIEVING THE DEVICE, THE WEB DETACHED IN THE VIA. RETURNING VIA WITH WEB INSIDE. A 6X3 WAS USED IN IT¿S PLACE. PATIENT DOING WELL. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19454713 ↗DeathApr 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT ADDRESSES THE SECOND WEB DEVICE. THE FIRST WEB IS CAPTURED UNDER MFR REPORT#: 2032493-2024-00267. IT WAS REPORTED THAT A PATIENT ENROLLED IN THE WEB PAS STUDY HAD AN ACOM ANEURYSM TREATED WITH A WEB DEVICE ON (B)(6) 2024. THE DEVICE DID NOT DETACH WITH THE USE OF THREE DETACHMENT CONTROLLERS AND WAS REMOVED. IT WAS REPORTED THAT DURING REMOVAL OF THE DEVICE, IT DETACHED WHEN IT WAS APPROXIMATELY 80-90% WITHIN THE CATHETER. THE DEVICE WAS REMOVED IN ITS ENTIRETY ALONG WITH THE CATHETER. THE COMPANY WAS ADVISED THAT THE CASE WAS CONTINUED WITH A SECOND WEB DEVICE. WHEN THE WEB WAS BEING ADVANCED OUT OF THE TIP OF THE MICROCATHETER, THE OPERATIVE RECORDS REFLECTED THAT IT MIGRATED BEYOND THE ROADMAP, CAUSING CONCERN FOR RUPTURE. THE RECORDS FURTHER STATE THAT THE WEB WAS STUCK IN THE WALL OF THE ANEURYSM, AND, WHEN REMOVED, DID NOT FILL THE ANEURYSM PROPERLY DUE TO A SHARP TURN IN THE ANEURYSM. THE WEB WAS REMOVED. IT WAS REPORTED THAT EXTRAVASATION AND SUBARACHNOID HEMORRHAGE WERE OBSERVED AND THE ANEURYSM WAS TREATED WITH EMBOLIZATION COIL IMPLANTS. WHEN EXTRAVASATION WAS CONTROLLED, THE PATIENT WAS SENT FOR A VENTRICULOSTOMY TO CONTROL INTRACRANIAL PRESSURES ATTRIBUTABL
Patient problems: Cerebral Edema; Extravasation; Intracranial Hemorrhage; Intracranial Pressure Increased; Ruptured Aneurysm
MAUDE report 19037082 ↗MalfunctionApr 3, 2024Material Separation; Separation Failure; Material Separation; Separation Failure
THIS REPORT ADDRESSES THE FIRST WEB DEVICE. THE SECOND WEB IS CAPTURED UNDER A SEPARATE REPORT. IT WAS REPORTED THAT A PATIENT ENROLLED IN THE WEB PAS STUDY HAD AN ACOM ANEURYSM TREATED WITH A WEB DEVICE ON (B)(6) 2024. THE DEVICE DID NOT DETACH WITH THE USE OF THREE DETACHMENT CONTROLLERS AND WAS REMOVED. IT WAS REPORTED THAT DURING REMOVAL OF THE DEVICE, IT DETACHED WHEN IT WAS APPROXIMATELY 80-90% WITHIN THE CATHETER. THE DEVICE WAS REMOVED IN ITS ENTIRETY ALONG WITH THE CATHETER. THE COMPANY WAS ADVISED THAT THE CASE WAS CONTINUED WITH A SECOND WEB DEVICE. WHEN THE WEB WAS BEING ADVANCED OUT OF THE TIP OF THE MICROCATHETER, THE OPERATIVE RECORDS REFLECTED THAT IT MIGRATED BEYOND THE ROADMAP, CAUSING CONCERN FOR RUPTURE. THE RECORDS FURTHER STATE THAT THE WEB WAS STUCK IN THE WALL OF THE ANEURYSM, AND, WHEN REMOVED, DID NOT FILL THE ANEURYSM PROPERLY DUE TO A SHARP TURN IN THE ANEURYSM. THE WEB WAS REMOVED. IT WAS REPORTED THAT EXTRAVASATION AND SUBARACHNOID HEMORRHAGE WERE OBSERVED AND THE ANEURYSM WAS TREATED WITH EMBOLIZATION COIL IMPLANTS. WHEN EXTRAVASATION WAS CONTROLLED, THE PATIENT WAS SENT FOR A VENTRICULOSTOMY TO CONTROL INTRACRANIAL PRESSURES ATTRIBUTABLE TO SUBARACHN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19036791 ↗InjuryNov 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN ELECTIVE PROCEDURE, THE WEB WAS DEPLOYED IN THE ANEURYSM, AND POST-DEPLOYMENT, A DYNA CT ENCROACHMENT INTO THE PARENT VESSEL WAS OBSERVED. THE PHYSICIAN MADE THE DECISION TO RETRACT THE WEB DEVICE; HOWEVER, THROMBUS WAS OBSERVED IN THE WEB AND RETRACTION OF THE DEVICE WAS DIFFICULT. AFTER THE WEB WAS RETRACTED, THE DECIDED WAS MADE NOT TO CONTINUE TREATMENT AT THIS TIME. THE PATIENT WAS NOT IMPACTED AND REPORTED TO BE DOING WELL, AND THE TREATMENT OPTION FOR A LATER DATE WITH A WEB OR STENT COIL TREATMENT IS UNDER CONSIDERATION.
Patient problems: Thrombosis/Thrombus
MAUDE report 18156820 ↗MalfunctionOct 30, 2023Difficult or Delayed Separation; Difficult or Delayed Separation
IT WAS REPORTED THAT THE WEB DEVICE DID NOT DETACH. THE WEB WAS THEN ATTEMPTED TO BE REMOVED FROM THE PATIENT. DURING RETRACTION, THE CONFIGURATION WIRE DETACHED FROM THE PUSHER LEAVING HALF OF THE DEVICE STUCK INSIDE THE CATHETER, AND THE OTHER HALF OF THE DEVICE STUCK OUTSIDE THE CATHETER. THE ENTIRE SYSTEM WAS ABLE TO BE REMOVED FROM THE PATIENT SUCCESSFULLY. THERE WAS NO PATIENT INJURY OR INTERVENTION. THE PATIENT IS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18031149 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.
- Z-0170-2023 · Sep 21, 2022 · Open, ClassifiedWEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Other Intrasaccular Flow Disruption Device devices
- WEB Detachment Controller WDC-2-MVIMicrovention, Inc. · Intrasaccular Flow Disruption Device1 recall0 reports
- WEB W5-4-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death23 reports
- WEB 17 SL 6x2 W5-6-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death20 reports
- WEB W5-5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device20 reports
- WEB W5-4.5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device19 reports
- WEB W5-7-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death16 reports
- WEB W5-6-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death15 reports
- WEB W5-5-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page